Class 1 Recall: GE Healthcare, LLC, Giraffe and Panda T-Piece Resuscitation Systems, and the Giraffe and Panda Bag and Mask Resuscitation Systems

The oxygen and air wall inlet fittings and/or labels on the back panel of the resuscitation systems may have been reversed during assembly. As a result, this could potentially reverse the air/oxygen concentrations; for

Mon Oct 15 2012By Other Author

The oxygen and air wall inlet fittings and/or labels on the back panel of the resuscitation systems may have been reversed during assembly. As a result, this could potentially reverse the air/oxygen concentrations; for example, a setting of 100% oxygen could have an output of 21% oxygen and vice versa. The settings of the blender knob would no longer be accurate.