Class 1 Recall: Respironics, Inc., Trilogy 100, 200, and 202 Ventilators

The Trilogy 100, 200, and 202 ventilators are being recalled because power supply components may be defective in some cases. If a ventilator’s power supply fails, the ventilator may unexpectedly stop functioning, thereby

Fri Apr 27 2012By Other Author

The Trilogy 100200, and 202 ventilators are being recalled because power supply components may be defective in some cases. If a ventilator’s power supply fails, the ventilator may unexpectedly stop functioning, thereby stopping ventilation. The ventilator alarm may also fail to sound. Ventilator power supply failures can result in serious adverse health consequences, including death.