Class 1 Recall Expansion: B. Braun Infusomat Space Infusion System
This is an expansion of a previous recall, initiated by B. Braun in January, 2012, due to the potential for breakage of the anti free flow clip catch located inside the infusion pump door. Breakage may occur when the IV set anti free flow clip catch
Fri Mar 23 2012
This is an expansion of a previous recall, initiated by B. Braun in January, 2012, due to the potential for breakage of the anti free flow clip catch located inside the infusion pump door. Breakage may occur when the IV set anti free flow clip catch is inserted improperly into the pump and the pump door is forced closed. Misloading of the anti free flow clip catch may create the potential for free flow of medication. Free flow, especially of narrow therapeutic range drugs, can cause life-threatening effects and injuries.