Class 2 Recall: Varian brand C-series Clinac, Trilogy and Novalis Tx, Software Versions 6.X and 7.X
The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch. Download
Sat Jan 08 2011
The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch.
Download