Class 1 Recall: Greatbatch Medical Recalls Standard Offset Cup Impactor Used for Hip Joint Replacement due to Inadequate Sterilization

Greatbatch Medical is recalling the Standard Offset Cup Impactor with a POM-C handle that failed sterility testing when sterilized in a dedicated instrument case.

Tue Dec 20 2016By Jonathan Payne

Class 1 Recall: Greatbatch Medical Recalls Standard Offset Cup Impactor Used for Hip Joint Replacement due to Inadequate Sterilization
 
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.
 
Recalled Product:
  • Standard Offset Cup Impactor with POM-C Handle
  • Model Numbers: T6318; T7821; T8042; T8043; T8044; T8088; T8143; T8177; T8184; T8333; T8468; T8487; T9196; T9316; T9348; T9360; T9747; T9894; T9954; T9955; T10243; T10281; T10287;T10600; T10491; T10753; T10861; T11209; T11340; T11506; T12230; T12699; T12861; T13344; T13480; T13642; T13722; T14384; T15311; T15752; T15949; T16427; T16611; T16661; T16934; T17203; T17238; T17321; T17650; T17703
  • Distribution Dates: July 30, 2004 to December 22, 2015
  • Manufacturing Dates: January 2004 to December 2013
  • Devices Recalled in the U.S.: 2906
Device Use
 
The Standard Offset Cup Impactors are reusable handheld devices used during hip joint replacement surgeries to implant cups in the hip socket (acetabulum). The device is provided as non-sterile and must be sterilized prior to use in surgery.
 
During hip replacement surgery, the ball (femoral head) is removed and replaced with a prosthetic ball, and the acetabulum is removed and replaced with a prosthetic cup.
 
Reason for Recall
 
Greatbatch Medical is recalling the Standard Offset Cup Impactor with a POM-C handle that failed sterility testing when sterilized in a dedicated instrument case. Non-sterile surgical devices can lead to infections, and other serious adverse health consequences, including death.
 
Who May be Affected
  • Health care providers using this device during hip replacement surgeries
  • All patient groups undergoing hip replacement procedures involving the Standard Offset Cup Impactor
 
What to Do
 
Greatbatch has developed new sterilization recommendations that meet acceptable sterility assurance levels. The device must be individually wrapped during sterilization processing.
 
On July 29, 2016, Greatbatch Medical informed its customers via letter to follow the updated sterilization instructions.
 
Contact Information
 
Customers with questions may contact Greatbatch via telephone at 1-619-498-9487 between the hours of 8:00 am (CT) and 5:00 pm (CT). Customers may also contact the company via email at FieldActionCenter@greatbatch.com.
 
Date Recall Initiated:
 
July 29, 2016
 
 
 