Valleylab™ FT10 Energy PlatformThe smart TissueFect™ tissue sensing technology in our Valleylab™ FT10 energy platform improves the speed and maintains consistency of all your hand-held devices by reading tissue composition in real time.4,5 So you can move through procedures more efficiently, for a better experience.
FORUMSView All (16)
1Reply3 months ago | maintenance service manual Equipment: Valleylab - FT10 Reply |
2Replies4 months ago | Device Correction - Errors E420 and E416 Covidien has issued a device correction to resolve these errors. As a part of the investigation into this issue, it was noted that upon insertion of a new (unused) LigaSure TM device, theValleylab TM FT10 Energy Platform running software versions 4.0.1, 4.0.2 and 4.0.3 may erroneously indicate that the LigaSureTM device was used previously. When this occurs, the energy platform will display error "E420 Usage Limit" or error "E416 Unknown Instrument," and the LigaSure TM device would not be allowed to be used. Upgrading the Valleylab TM FT10 Energy Platform to the newly released software version 4.0.4 will eliminate this issue. Software will be provided by Covidien at no charge. Reply |
0Replies4 months ago | E233 after change power supply FT10 IT come with bead power supply . I think it will be easy to do. So I bought used from ebay. But after installing I have E233. Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Electrosurgical Unit (ESU) Companies
VIDEOS
FEATURES
- Faster vessel-sealing speed with our LigaSure™ instruments
- Improved TissueFect™ tissue sensing technology
- An even easier to use, intuitive system interface
NEWSView All
Popular Biomedical Training Instructor Launches New Business
Quality training is essential for the Biomedical profession, but it's often hard to find. BC Group is proud to endorse and recommend this new and innovative option for our industry.