Sorvall - Evolution RC
by Sorvall

Follow Us
DESCRIPTION

Employing leading-edge technology, the Evolution RC offers unmatched productivity with a six liter capacity, speeds to 26,000 rpm, and simplified sample handling.

The Evolution RC offers shorter processing times with more throughput – and of course, maximum reliability.

The unique DuoTherm system provides continuous, alternate heating and cooling to hold temperatures to within 2°C, even in standby. This not only prevents samples from freezing and chamber frost, but also eliminates frequent starting and stopping of the compressor – thereby extending compressor life.

FORUMSView All (7)
Ask a New Question

2

Replies
-scientific_1916
3 years ago
Zaphod B. Scientific-Laboratory electronics engineer NZ
Anyone ever had a motor drive assy issue whereby there is a motor enable and a run signal reaching the first pcb yet the lower drive module "enable" red LED is not turning on? The error message comes up in the display as "Tach signal not present" but this is only because the motor drive control module is not being "enabled" I'm guessing the fault s/w table isn't smart enough to isolate the fault any further? Don't want to have to purchase a new motor drive module as this is probably an smd logic ic fault on one of the motor drive pcb's which can easily be repaired. Any help will be greatfully reciprocated thanks Z Reply

7

Replies
-olNick
4 years ago
sensor offset error does not clear with Clear key
Manual says this is a minor error and to simply clear it. The Clear key does not clear it. Please help. The machine was fine otherwize.Reply

8

Replies
-raulasoto
7 years ago
Power Fail error

when I turn on the main switch on the monitor appear POWER FAIL error msg I need to do whant can I do to resolve this alarm thanks att raul

Reply
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Centrifuge Companies

FEATURES

Fastest, easiest processing in the industry

  • With the F8S-6x100Y SUPER-LITE composite rotor, users will process 6 liters of sample in a single run. Combining speeds to 26,000 rpm with faster accel/decel times than competitive centrifuges, the Evolution RC reduces overall spin times and optimizes productivity.

Rotors and Accessories

  • Sorvall's complete line of rotors, tubes and bottles will meet all application needs. With the addition of the newly designed 1L bottle, offers users the complete system – centrifuge, rotor and bottles all in one.

Consistent performance, longer life

  • The built-in Constant Power System (CPS) provides power factor correction to ensure peak performance even under poor power supply conditions. Input voltage can vary from 187 – 253 V without speed or temperature control degradation.

Easily accessible, hands-free rotor access

  • The Evolution RC’s deck is only 850 mm (33 inches) from the ground – the lowest profile on the market – which makes the rotor chamber and control panel easily accessible.

SPECIFICATIONS

Height1295 in
Length 1055 in
Weight860 lbs
Maximum Speed26000 rpm
Width710 in

Additional Specifications

         
    Drive: Direct Brushless, DC Motor with DuraFlex™ gyro suspension     
    Accel/decel rates: 3 acceleration, 4 deceleration     
    Speed control accuracy: +0.2% or 20 rpm, whichever is greater     
    Temperature control: +2 to +40 °C     
    Temperature accuracy: +2 °C     
    Ambient temperature range: 10 - 38 °C     

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree