B. Braun - Outlook 100
by B. Braun

Follow Us
DESCRIPTION
It lets you plan for the future today.

The Outlook Safety Infusion System helps to ensure that the right patient receives the right medication in the right dose from an authorized clinician at point-of-care. The Outlook 100 incorporates dose limits to help reduce human errors in medication administration.

FORUMSView All (29)
Ask a New Question

1

Reply
-Ccortez
5 years ago
Electrical schematics
Does anyone know where to find the schematic for this device. TiaReply

4

Replies
-Acela Medical
5 years ago
Go-No/Go Gauge
Does anyone know where I can purchase the Go-No/Go Gauge or the dimensions of the gauge for the Yoke Adjustment Test?Reply

0

Replies
-Stan Skaggs
7 years ago
PM notes to keep in mind
Batteries in this unit are NiCD alarm batteries which I found always are dead. The lithium memory batteries last forever evidently (9 years+), and main batteries a couple of years or so depending if the nursing staff will plug them in. The petal covers wear thin and have to be replaced annually for me (every other PM cycle). I check the gaps with the gap tool and found not likely to be off, but best to check the gap. I hate changing the alarm batteries and grimace when I pull one out because it has nine leads that all have to be unsoldered on tiny pads through the board. Why did they use NiCD batteries for goodness sakes?? The green boards are easier than the blue ones to PM and change batteries with.Reply
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Infusion Pump Companies

FEATURES

  • Promotes patient safety by helping to avoid accidental under- or overdosing using audible and visual alerts
  • Provides flexibility for each care area—ability to customize dose limits to your hospital drug formulary
  • Promotes clinician choice with dose limit override when limits are exceeded during manual programming and titration
  • Visually indicates dose selection with DoseGuard software
  • Easy to learn
  • Manual operation RATE, VOLUME, RUN
  • Automatic Dose Rate Calculator
  • Large, highly visible keys
  • Reduce nuisance air-in-line alarms with air-trapping cassette
  • Stack-able modular design reduces bedside cord clutter and errors due to channel confusion
  • Comprehensive line of needle-free administration sets including UV light resistant tubing that protects drug efficacy

SPECIFICATIONS

Volume to be Infused0.1 to 999.9 ml
Height4.5 in
Length 8.75 in
Flow Rate Range0.1 to 999.9 ml/h
KVO Rate3 ml/h or selected rate whichever is lower
Weight5 kg
Flow Accuracy5 %
Infusion Pump TypeGeneral-Purpose
Width12 in

Additional Specifications

Power: 120 VAC

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.