FORUMSView All (5)
2Replies7 years ago | slice count how could I know the slice count and other information helpping me to know the condinoin of a used machine. Reply |
7Replies8 years ago | no scans with xray I got the Host critical error message #1 even when the system boots up fine. xray tube, ebox and hv tank have been replaced already. I can run scans with the xray off. anybody has an idea on what is happening? Reply |
0Replies10 years ago | Communications Problems Acq board never light the comm_en led. When I check the IRS Log in Procmon an error shows up: "AQC : Still waiting to RISC getting to ready state machine". I've change fiber optic cable, ACU, MCU Receiver, Transmitter and no change. Anyone can help me. Thanks. Reply |