In exams where image quality is paramount, the Mx8000 Quad provides high-resolution images with Isotropic Imaging capability. Its true 0.5 mm slice thickness delivers excellent image quality for critical areas such as I.A.C.s, P.E. studies, and joint examinations.

FORUMSView All (5)
Ask a New Question

2

Replies
-tough
7 years ago
slice count

how could I know the slice count and other information helpping me to know the condinoin of a used machine.

Reply

7

Replies
-tonyball
8 years ago
no scans with xray

I got the Host critical error message #1 even when the system boots up fine. xray tube, ebox and hv tank have been replaced already. I can run scans with the xray off.

anybody has an idea on what is happening?

Reply

0

Replies
-ICONPR
10 years ago
Communications Problems

Acq board never light the comm_en led. When I check the IRS Log in Procmon an error shows up: "AQC : Still waiting to RISC getting to ready state machine". I've change fiber optic cable, ACU, MCU Receiver, Transmitter and no change. Anyone can help me. Thanks.