When you need to get to a patient fast, you need a durable, mobile X-ray imaging system. GE Healthcare’s AMX-4+ analog X-ray system answers that challenge by providing high-performance in a compact, easy-to-maneuver package. The rotating arm and tube simplify positioning and facilitate bedside studies, while the maintenance-free battery produces up to 50 high-quality exposures with a single charge.  

FORUMSView All (166)
Ask a New Question

2

Replies
-Jere
2 months ago
Starts to Rotor then shuts down.

Hello,

To start at the beginning ... 1st) Tube/collimator was tough to move up/down to column.  Replaced upper brake to column, brake wires shorted to housing when re-assembling.  Repaired wires, routed wires a different way, re-assembled ... Tube/collimator easy to move up/down.  Charged portable overnight, next day went to verify system working before giving to the techs .... rotor cuts out in 2 seconds & system restarts, tests passed, no errors, CB never trips off.   

2nd) Now AMX-4 + when you rotor from Hand switch, can hear relays click, 2 seconds later the displays go blank and startup test pass, display is ready again.   I can hear the rotor spinning in tube,  tube resistance is good. I verified capacitor values on right side. All fuses are good. Reseated Relays on circuit boards.  I have a replaced 1 board at time to the following .... Display Bd, CPU, Rotor Bd, FIL/Kvp Bd.  No errors appear after startup tests.  Can't do any cals because rotor doesn't last but 2 seconds.    Does anyone have any other idea's? 

Thanks

Jeremiah

Equipment: GE HealthCare - AMX 4 Plus

Reply

0

Replies
-naveedbhatti44
3 months ago
Error 431 & 433

I have facing problem when exposure button is pressed error papers 431 & 433 any help to resolvsolve this

Reply

1

Reply
-Faisal005
9 months ago
Test4 failed

Turned on AMX4 but test4 failed please help how to solve it 

Equipment: Amx4

Reply
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Mobile Radiography Companies

FEATURES

[list] [*]Dual drive motors and oversize casters facilitate movement; even in taxing environments. [*]Ease of use, effortless maneuverability and 270 degree column rotation add positioning flexibility across diverse care areas and exams, even in tight spaces. [*]Large installed base results in operator familiarity, reducing training costs. [/list]

SPECIFICATIONS

Maximum Focal Spot0.75 mm
Height70 in
Length 45 in
Weight1080 lbs
Width25 in

Additional Specifications

Remote x-ray control: Handswitch-Standard

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree