It starts with the imaging chain, which features a 12-bit CCD-based camera.Then there’s AutoEx, which automatically adjusts key image-acquisition parameters in real time for optimum quality. Meanwhile, closed-loop Automatic Brightness Control and patented extended Dynamic Range circuitry maintain contrast and minimize blooming.

FORUMSView All (27)
Ask a New Question

0

Replies
-VMGT
a year ago
Generator error 3601163
Generator error 3601163 is being triggered everytime I try to use the large focal spot. I already measured resistance and continuity directly into the tube and also in the generator side to make sure cables are ok as well. They are just fine. I also did what the manual says, I replaced the heater board and it is still the same. Does anyone have a recommendation?Reply

0

Replies
-RDK
2 years ago
table angulation
Have an isse with table angulation at zero deg reference, not consistant, motr, pot, board and MRAS controller all have been changed out, loaded software and backups. any ideas?

Equipment: GE Healthcare - Precision 500 D

Reply

0

Replies
-RDK
2 years ago
Precision 500 angulation zero reference problem
I have an issue with A GE prcision 500 table, the table has zero references that change (ex. 0 deg ref to start, if i power th room down or reset it may be at -5 deg or another number there is no consistancy), motor pot,board, MARS system controller have all been replaced, all software reloaded and table calibrated multiple times. Any ideas?Reply
DOCUMENTS / MANUALSView All

FEATURES


  1. Ergonomic SMART Handle table positioning and tube controls enhance productivity and patient comfort.
  2. Automated dose and image management tools reduce radiation to help protect both patient and technician.
  3. 12-bit CCD-based camera, 1024 x 1024 imaging matrix and the Fluoro Noise Reduction (FNR) algorithm combine to create an imaging chain that produces images of extraordinary clarity.

SPECIFICATIONS

Maximum Focal Spot73.5 in
Height34.8 in
Length 83.9 in
X-ray Generator Power Rating80 Kw
Motorized Tilting/Pivot Rotation90
Width28.3 in
This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.