Hospira - Plum A+3
by Hospira

Follow Us
DESCRIPTION
The Plum A+3 general purpose infusion system remains the proven choice for caregivers and clinicians alike

With more than 325,000 pumps installed worldwide, the Plum A+3 general purpose infusion system remains the proven choice for caregivers and clinicians alike. The scalable platform can be initially deployed, and later upgraded to include Hospira MedNet™ safety software to meet your evolving patient care needs. The unique, innovative PlumSet™ cassette technology and time-saving features simplify medication delivery and help to enhance safety for patients, clinicians and caregivers.

FORUMSView All (6)
Ask a New Question

0

Replies
-SirSeacow
a year ago
CE Board Passwords
Is there a default password to get into the CE board, ive tried admin, admin... admin, 4552401... neither work.Reply

0

Replies
-SirSeacow
a year ago
CE Board reset / passwords
I'm trying to get into the CE Board settings, I've tried all the passwords I know, can anyone start a list of a default password to get into the boards?Reply

4

Replies
-BioMedTech631
2 years ago
A+ mechanism
my facility has been running the Abbott Plum A+ pumps for a while , trying to keep them going for the next or so until we upgrade to 360s. We've had several mechanism failures recently, most due to pressure out of cal. Are there any companies that repair these mechanisms instead of us buying new mechs?

Equipment: Hospira - A+

Reply
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Infusion Pump Companies

FEATURES

[list]  [*] Deliver infusions with confidence with the unique PlumSet™ technology                  [*] Automated second-line delivery — eliminates the need to raise and lower infusion containers, improving ease-of-use and efficiency                  [*] Concurrent delivery of 2 medications through a single patient line                  [*] Air-in-Line detection and elimination — air trap captures up to 2 mLs of air                  [*] Accumulated air can be removed by automated back-priming without disconnecting the patient line, reducing the risk of contamination                  [*] Deliver secondary infusions via IV container or syringe                  [*] Reliable performance, steady, consistent and proven                   [*] One integrated system works across the entire spectrum of clinical care                  [*] Single channel (2 concurrent medications) or triple channel (6 concurrent medications) models available  [/list]

SPECIFICATIONS

Volume to be Infused0.1 to 99.9 ml
Height19 in
Length 14 in
Weight28 lbs
Infusion Pump TypeGeneral-Purpose
Width15 in

Additional Specifications

Power Requirements: 120 V, 50-60 Hz, 150 VA  Delivery Rate Accuracy: ± 5% (1-999 mL/hr), ± 10% (0.1-0.9 mL/hr)

FDA AlertsView All

Class 1 Recall: Hospira Plum A+ and Plum A+3 Infusion Systems...

The Plum A+ and A+3 infusion pumps have an alarm that should sound when a therapy is interrupted. Some of the alarms may fail to sound in situations that should trigger it.