DESCRIPTION
Mobile, electrohydraulically operated surgical table designed to support all general surgical procedures

The Steris AMSCO 3085 and 3085 SP surgical tables are among the most well-designed general surgery tables on the market. Although they are being replaced with newer, more effective tables, due to Steris's implementation of the FDA's automatic upgrade for patient weight capacity, all Steris 3085 SP tables (from any manufacture date) are now approved by the FDA to hold patients up to 1,000 Lbs. in certain positions.    
This gives the 3085 and 3085 SP unprecedented versatility and flexibility in patient positioning for every surgical specialty.   
And with the optional ACT (Advanced Control Technology) interface, the AMSCO 3085 can be networked to various Operating Room Control Systems (ORCS). This provides surgeons with direct control of the AMSCO 3085 SP using verbal commands or via touch screen.

FORUMSView All (11)
Ask a New Question

1

Reply
-Dr ZZ
5 months ago
Need Service Manual Amsco Surgical Table

Does anyone have a service manual for Amsco Surgical Table model BL57727-200?

Leg sections not raising correctly. 


We get some rise on leg sections but then it hits a point where it starts tilting main table back. 

Faulty hydraulic tube or limiter switch maybe??

Can email manual to:

Service@TurnerMedicalRepair.com


Equipment: Amsco surgical table

Reply

0

Replies
-mitchell earl
2 years ago
Steris # BL57727-330
Hand control blinks but no locking action. On bed controls do not work either. They changed controllers and power cord with no results. Anybody have a service manual they can send me? Any ideas as to the root cause of this issue? bpsalabama@gmail.com Thank you for your help in advance.Reply

25

Replies
-Robert Sulenov
4 years ago
remote stopped working
I need help determining the following problem. Our remote suddenly stopped working. It started with all lights would flash when you first turn it on and then everything went out. Sometimes it would turn on normally and you could use it, but only for a few seconds and then it dies. After that it makes some time of a clicking sound when you press ON and OFF, but you can't control anything. This is my first encounter with this technology and any help/advice is greatly appreciated. Thank you, RobertReply
SERVICE COMPANIESView All Surgical Table Companies

SPECIFICATIONS

Height27 in
Weight737 lbs
Tredelenburg 25 degrees
Patient Weight Capacity1000 lbs
Width14.5 in

Additional Specifications

Electricity: 100 V, 50/60 Hz, 5.0 Amp; 120 V, 50/60 Hz, 4.5 Amp;     
220 V, 50/60 Hz, 3.0 Amp; or 230/240 V, 50/60 Hz, 2.5 Amp.     
     
Reverse Trendelenburg – 25° maximum from horizontal.     
Height: 27" (686 mm) minimum to 44" (1118 mm)     
maximum.     
Side Tilt: 18° maximum to right or to left of horizontal.     
Back : up 55° maximum (80° in REVERSE orientation) or     
down 25° maximum (105° in REVERSE orientation) from horizontal.     
Leg: up 80° maximum (55° in REVERSE orientation) or     
down 105° maximum (25° in REVERSE orientation) from seat section.     
Flex: back section down 20° maximum with seat section down 25° maximum from horizontal.     
Reflex: back section up 25° maximum with seat section     
up 35° maximum from horizontal.  