Rigel Medical - SafeTest 50
by Rigel Medical

Follow Us
DESCRIPTION

Analyzer Measurement & Simulator 

The Rigel SafeTest 50 is a rugged, reliable and dedicated medical safety analyzer. It is ideal for testing high volumes of medical and laboratory equipment that does not need patient lead testing.    With a large color display and a color-coded user interface, its easy to select the required tests with a single key press. A fast selection of the test routine makes the testing process speedy and dependable.    While compact, the SafeTest 50 includes a range of safety tests to enable compliance with a range of international safety standards, including leakage testing to NFPA-99, ES 60601, IEC 62353 and 61010, ground bond testing to NFPA-99, IEC 62353 and 61010.    Full manual control offers the benefit of executing only those specific tests that are required and provide the user with full control of the power cycles, making testing simple, easy and fast.

FORUMSView All (2)
Ask a New Question

8

Replies
-hitechluddite
8 years ago
Good unit?

Looking to purchase a line leakage/safety test meter and was leaning toward the BC biomedical SA-2001 then the Safe Test 50 came along. They are at the same price point which would be the better choice? It will be used in conjunction with PM's on Tissue processors. Stainers, Coverslipper, embedders etc.

Reply

0

Replies
-Data Entry Coordinator
8 years ago
Rigel Medical SafeTest 50 Electrical Safety Analyzer - USA

The Rigel SafeTest 50 is a rugged, reliable and dedicated medical safety analyzer. It is ideal for testing high volumes of medical and laboratory equipment that does not need patient lead testing.

With a large color display and a color-coded user interface, it’s easy to select the required tests with a single key press. A fast selection of the test routine makes the testing process speedy and dependable.

While compact, the SafeTest 50 includes a range of safety tests to enable compliance with a range of international safety standards, including leakage testing to NFPA-99, ES 60601, IEC 62353 and 61010, ground bond testing to NFPA-99, IEC 62353 and 61010.

Full manual control offers the benefit of executing only those specific tests that are required and provide the user with full control of the power cycles, making testing simple, easy and fast.

Reply
DOCUMENTS / MANUALSView All

VIDEOS

Rigel Medical SafeTest 50 Electrical Safety Analyzer - USA

FEATURES

  • Rugged, reliable and portable design
  • Large color display and easy to use interface
  • Tests and fault conditions can be selected with a single key press
  • Leakage testing to NFPA-99, ES 60601, IEC 62353 and 61010
  • Ground bond testing to NFPA-99, IEC 62353 and 61010
  • Accurate high current, low energy earth bond testing

Additional Specifications

  • Application    L-N, L-E, N-E and touch voltage (IEC 61010)
  • Range    0.0V – 300VAC
  • Resolution    0.1Hz
  • Accuracy    ± 2% ± 5 digits (between 10V - 270VAC)
  • Mains Frequency    45 to 66Hz