Class 1 Recall: Abbott Diabetes Care, FreeStyle InsuLinx Blood Glucose Meters

The FreeStyle InsuLinx Blood Glucose Meter measures sugar (glucose) in blood drawn from the fingertips of people with diabetes to monitor blood sugar levels.

Mon Apr 15 2013By Kaylee McCaffrey

 

Reason for Recall:

Incorrect Test Results at Extremely High Blood Glucose Levels.

At extremely high blood glucose levels of 1024 mg/dL and above, the FreeStyle InsuLinx Blood Glucose Meter will display and store in memory an incorrect test result that is 1024 mg/dL below the measured result.

For example, at a blood glucose value of 1066 mg/dL, the meter will display and store a value of 42 mg/dL (1066 mg/dL – 1024 mg/dL = 42 mg/dL).

Blood glucose levels at 1024 mg/dL and above are very rare. However, if high blood glucose levels of 1024 mg/dL and above occur, they are a serious health risk that requires immediate medical attention. If the FreeStyle InsuLinx Blood Glucose Meter displays an inaccurate low result, there may be a delay in identification and treatment of severe high blood glucose levels (hyperglycemia), or inappropriate treatment may be administered to the patient. This could lead to serious injury or death.

 