Reply
-AaronGil
a year ago
Error: 351.6760

Two weeks ago I received this device from a customer. I pulled this error code from the logs and found the error definition: "The calibration flag is not set. The module is not calibrated." I preformed a calibration on the device and returned the device to service. The device is now back today with the same issue. 

Reply

2

Replies
-me_samm
3 years ago
How to enable profiles
When connected to the PCU and selecting the channel, I get a screen that says "Profiles have been disabled. Have PC Unit serviced by qualified personnel." How do I enable the profile? ThanksReply

3

Replies
-DCRUMP2016
5 years ago
Alaris PCA module 8120
Hi I thought there was a way to run a volume test on the Alaris PCA 8120 module but I haven't been able to fine it. If anyone knows of a way to do this please share. Thanks Reply