The Force FX™ generator C with Instant Response™ technology is designed for the most demanding monopolar and bipolar procedures, with a PER of 98 in cut. It adjusts automatically, responding to tissue changes, maintaining power delivery, and minimizing drag. This offers a smoother cut through all tissue types.

FORUMSView All (47)
Ask a New Question

0

Replies
-ferideatasoy
2 days ago
161 hatasi
Cut islemi esnasinda 161 hatasi aliyorum. Roleyi çikarinca da cut islemi uzun sürünce 161 hatasi almaya devam ediyorum

Equipment: Valleylab - Force FX-C

Reply

7

Replies
-AngelCtrz
4 months ago
Error code 16 and some other issues.

I have this problem and already read the service manual but doesn't have the transformer or inductive resistors. I have the fluke 454A but doesn't have the 30 ohms resistant option, with 100 and 500 would be enough? The error 16 starts when put on the plate on the patient. 

Also read in the forum that could be "Scaling relays in the control board", how can I probe them? I doesn't have another Force Fx.

Some days ago I made the REM calibration because the equipment was showing green REM alarm when plug in the plate even without patient. Also the trouble that made the hospital call me was that the equipment wasn't turning on right, it had a problem in the initial test.  When the lights were on bipolar it started as usual but were restarting the autotest when passing trough cut. 

Should I try to fix It or is better just change the equipment with all that issues?

Appreciate any help 

Equipment: Valleylab - Force FX-C

Reply

0

Replies
-Gladki
7 months ago
ERROR 212/217
I have a problem, my force fx displays errors 212/217 alternately. There is no such error as 217 in the service manual. Does anyone know what it means? In the device in service mode, there are 10 steps, but in the service manual there are only 8 steps described. How to calibrate step 9 and 10?

Equipment: COVIDIEN FORCE FXc

Reply
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Electrosurgical Unit (ESU) Companies

FEATURES

  • Power delivered remains virtually constant, regardless of tissue type
  • Improved performance at lower power setting minimizes the risk of tissue damage and neuromuscular stimulation
  • Three internal microcontrollers reduce system reaction time and increase the system's processing speed

SPECIFICATIONS

Height4.38 in
Length 17 in
Weight18 lbs
Width14 in
NEWSView All
Popular Biomedical Training Instructor Launches New Business

Popular Biomedical Training Instructor Launches New Business

Quality training is essential for the Biomedical profession, but it's often hard to find. BC Group is proud to endorse and recommend this new and innovative option for our industry.

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree