Instrumentation Laboratory - GEM Premier 4000
by Instrumentation Laboratory

Follow Us
DESCRIPTION

Blood Gas/ pH Analyzer 

The GEM Premier 4000 is the most comprehensive whole blood analyzer available today, offering a flexible analyte menu including pH, pO2, pCO2, electrolytes, metabolites and onboard CO-Oximetry. Yet it’s remarkably easy to use and maintenance-free, making the GEM Premier 4000 ideal for busy labs, use at the point-of-care and everywhere fast turnaround is mandatory. Easy to learn, simple to operate and complete, IL’s GEM Premier 4000 with iQM revolutionizes whole blood testing.

FORUMSView All (17)
Ask a New Question

0

Replies
-Phillip L. Lyons
a year ago
Is PHI saved.
Does, the GEM Premier 4000 save PHI on it. I need to confirm that no PHI is on the device before having the equipment submitted for destruction.

Equipment: Instrumentation Laboratory - GEM Premier 4000

Reply

4

Replies
-Raynorshine Enterprises
7 years ago
Password?
I have a few of these for which I need to figure the stock password protocol or default. Any suggestions?Reply

0

Replies
-matar hnh
8 years ago
gem4000

error code 230 what does it mean thank you in advance

Reply
DOCUMENTS / MANUALSView All

FEATURES

  • Onboard CO-Oximetry and a full complement of parameters meet every testing need
  • Self-contained, multi-use cartridge PAKs include all testing components
  • Maintenance-free – just replace the cartridge every 30 days
  • iQM, IL's proprietary Intelligent Quality Management system, automates quality control
  • GEMweb® Plus, IL's unique remote management software, enables access to any networked analyzer to see results, monitor performance and assure standard configurations

SPECIFICATIONS

Height18 in
Length 15 in
Weight44
Width12 in
NEWSView All

Device Alert

Instrumentation Laboratory GmbH has issued an information that the package inserts for certain product lots of GEM Check Coag Whole Blood Controls contain incorrect data. Instrumentation Laboratory—GEM Check Coag Whole Blood Controls: Package Insert May Contain Incorrect Acceptable Performance Range See attached PDF under Manuals & Resource ...

FDA AlertsView All

Class 1 Recall - GEM Premier 4000 PAK Cartridges for Use on the...

Potassium test results on the GEM Premier 4000 are too low when compared to a reference analyzer; with biases exceeding the allowable error claim of plus or minus 0.5 mmol/L by as much as 2.0 mmol/L. Use of this product may lead to

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.