Philips - V24C
by Philips

Follow Us
DESCRIPTION

Muliti-Parameter 

Compatible with Hewlett Packard CMS, Agilent CMS and Philips CMS modules.

FORUMSView All (1)
Ask a New Question

0

Replies
-Codytj1973
5 years ago
PM template example
Initial Inspection: The unit has no physical damage. The unit will need to have the PM Checks completed. PM Checks: [X] Verify Charge circuit [X] If applicable verify printer operation [X] Control and Switches- Verify Operation [X] Indicators and Displays- Check Operation [X] Touch Screen Calibration [X] NIBP Read Back [X] Respiratory Read Back [X] SPO2 Read Back ESI Results: POWER CORD RESISTANCE 0.150 OMHS LEAKAGE TESTS, EQUIPMENT PWR ON CASE INT, NORM POL, CLSD GND, CLSD NEU 0.1 uA RMS CASE INT, NORM POL, OPEN GND, CLSD NEU 117.0 uA RMS EQUIPMENT CURRENT 0.52 A RMS COMMENTS: PASSED ESI TEST USER TIME: 0.10 HOURS Unit Status: The unit passed the PM test. The work order is completed. Reply