Volume Navigation helps to increase diagnostic confidence by combining multi-modality imaging with any ultrasound mode, including Contrast Enhanced Ultrasound and CFM. LOGIQ E9 knows the next step of a scan and helps users get there like no other ultrasound system can. It’s all part of the technologically advanced Scan Assistant, customizable scanning protocol. Designed by users, Scan Assistant enables users to take more control over the important elements of an exam by doing the little things for users. That’s because it automatically inserts comments, initiates and autocompletes required measurements, and much more. Scan Assistant does things that an ultrasound system has never done before, including automatically steering color Doppler and setting up imaging controls and modes for users. Now users can truly concentrate less on keystrokes and more on patient care.

FORUMSView All (61)
Ask a New Question

1

Reply
-KBarr52
8 months ago
System Error code

Hi All, 

I PM'd a system that was freezing on startup (B4)- main power supply had a rug of dirt on the fans and it appeared to resolve the issue.  Worked fine until a system error message appeared on startup, after reboot the alphanumeric keyboard is not connecting.  Would anyone know if I should stay in the direction of BEP for troubleshooting,  or could this be a separate operating panel issue?  Any guidance is appreciated. :)

Equipment: GE Logiq E9

Reply

0

Replies
-ESTDK
10 months ago
CMS password needed to perform DRX DC Offset calibration

We have 2 of these machines that were donated and sent missing the hard drives, so we reload the software on a hard drive and everything works except it says DRX DC offset calibration is required.  Looking in the service manual it is asking for a CMS password.  Tried contacting GE but because we did not buy from them but these were donated to the out of USA hospital they will not assist.  Is any additional information needed for someone to get us the CMS password so we can do this calibration and make these machines functional, or are these destined to the scrap pile of other improperly donated equipment to these low income countries.  Thank you.

Reply

2

Replies
-Medifixer
2 years ago
GE Logiq 9 "no output device configured"
Hello, I am trying to get Dicom set up. Retrieve worklist and send images to PACS. I believe I have some incorrect entries in my Dataflow but I have not figured out what it is. Please someone tell me why am getting these errors: MSG_LOCAL_WORKFLOW_FAIL and "No output devices configured. Images are still in clipboard" Your advice will be greatly appreciated.Reply
SERVICE COMPANIESView All Ultrasound Systems Companies

SPECIFICATIONS

Cart Based1
Height45.3 in
Length 32.7 in
Weight298 lbs
Clinical ApplicationsColorectal
Imaging Capabilities2D/3D
Color-flow Mapping1
Doppler Modes3-D, Color, Continuous wave (CW), Pulsed wave (PW), Power Doppler
Image Display19 in Color Flat Panel
Width23 in
NEWSView All

2011 Best in KLAS Awards for Medical Equipment and...

New KLAS Report Includes Inaugural Infrastructure Market Research > Orem, UT – June 15, 2011 – KLAS today announced the release of their highly anticipated 2011 Top 20 Best in KLAS Awards: Medical Equipment & Infrastructure report. The coveted awards are based solely on data from customer surveys of hospital and clinic executives, administrator...

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree