They offer healthcare providers faster data availability, better quality control, increased regulatory compliance and improved administration. HemoCue 201 DM systems allow healthcare providers to ensure the authorized and appropriate use of HemoCue analyzers. By means of a wide range of settings and definable options, the systems can be made to prompt the operator for identification, lot numbers, quality control and other required information during analysis. This data is then transmitted to clinical information systems, which leads to improved staff and patient management and to more effective billing.
1ReplyMon Jun 13 2022 | Error code E02 While cleaning optics glass in camera fall over of device. I accurate disassemble device, disassemble camera, glued glass in place. After assebling back get code E02 (low lever of measure diode). If get out ligth transfer colimator get error E01 (low reference diode). When I try to run with no covers get error E03(high black measure). Is any specific while assebling it back? While dasassemble nothing fall of device. When assemble IR filter place over green diode, is that correct?Reply |
0RepliesTue Sep 11 2012 | lost pass code we hire a new teck and we lost the code to use the machine can you help we have a hemocue 201 dm thanks Reply |