0RepliesSat Apr 17 2021 | Technical Support Can anyone help us, please provide us a service manual of BB-100 Incubator, we really need data/specs for our calibration. Thank you very much. JoaquinReply |
5RepliesFri Mar 19 2021 | Actual temp on chamber is lower by 1.8 deg vs the set point. chamber temp is 34.2degC lower by 1.8deg C versus the set point of 36degC. There are two(2) potentiometer on the control board (1009BB100). Can somebody share the User Manual and circuoit diagram of the control board (1009BB100).Reply |