Bioquell - Microflow Class II
by Bioquell


Washer/Decontaminator
The Microflow Class II Advanced Biological Safety Cabinet provides operator and product protection for work with up to category 3 pathogens:
This cabinet has been developed specifically to utilize Hydrogen Peroxide technology for bio-decontamination, offering a realistic alternative to formaldehyde fumigation. The cabinet operates in conjunction with the Clarus™ L vapour generator to provide fast, efficient, validatable and safe gassing of the cabinet. Specifically configured for use with Hydrogen Peroxide, all construction materials have been selected for compatibility.
During bio-decontamination the cabinet operates at negative pressure eliminating leaks to the atmosphere, gas passes through all filters to ensure efficient removal of all contaminating particles. Shorter cycle times for bio-decontamination means minimal downtime in the laboratory and all units feature an anti-blowback valve allowing bio-decontamination to take place without closing the exhaust dampers.
2RepliesSat Dec 18 2021 | Battery fail message. Please help. Good afternoon, I need help urgently. The bio security cabinet Bioquell Microflow Class 2 displays the following message: Fan 100%, Battery Fail. I found an open fan fuse and already replaced it, now the fan works as soon as it is connected to power. I think the internal battery on the main board will need to be replaced. I don't know if I am correct and if after that I should go into some mode to clean up the error. Thank you. Reply |