Rayto - RT-2100C
by Rayto


Microplate Reader
The RT-2100C microplate reader offers a 96-well plate with the ability to perform multiple tests in a single plate. Other features include large LCD display with touch screen, and adjustable plate shaking, speed and time.
1ReplyFri Sep 20 2024 | Service manual Need service manual of RT2100C rayto microplate reader Equipment: RT2100 C Reply |
3RepliesThu Apr 14 2022 | Service manual RT-2100C Microplate Reader|Rayto service manual ?Reply |
-- Fully automatic
-- Easy Windows operation system with touch screen or mouse, large LCD display 8-channel optical fiber system enables 5 secs' reading for the whole plate
-- Random positive and negative control setting
-- Multi-assay enables up to 12 different assays on one plate
-- Large memory, up to 100 test protocols and 10,000 can be stored
-- Powerful QC function: Grubs, Westguard Multi-rule, Levey jennings plot (optional)
-- Plate shaking
-- Auto power-on self-check
-- Additional lamp saving
-- Multiform result output including patient comprehensive report
-- Ability to communicate with PC for data management (optional)
Absorbance Range: 0 - 4.000Abs
Resolution: 0.001Abs(displayed),0.0001Abs(calculated)
Accuracy: 0.1% or 0.005Abs
Type of Microplate: Standard with 96-well or other kind of microplate and strip
Light Source: Halogen lamp, 8V/50W
Wavelengths: 405, 450, 492, 630nm, 4 more filters optional
Wavelength Accuracy: 1nm
Band width: 8nm
Calculation Method: ABS, %ABS, Cut-Off, Single Standard, --Curve, Multi-Percent, Percent Log, Linear, Exponential, Power, 4PL Regression
Reading Speed: 5 seconds for 96 well plate (single wavelength)
Shaking Plate: Shaking time and speed adjustable
Memory: More than 100 programs, up to 100,000 test results
Printer: All printers compatible with Windows
Interface: RS-232, PSTN (optional)
Display: 6" LCD
Input: Touch panel or mouse
Output: External printer (optional)
Power Supply: AC 110V - 220V , 50-60Hz
Net weight: 8Kg
Dimension: 460*330*200