GE HealthCare - BrightSpeed Series
by GE HealthCare

Follow Us
DESCRIPTION
Capable. Compact. Convenient.

The BrightSpeed Series gives you full multi detector CT capabilities in three smaller systems that are as easy to select and site as they are to operate and own.

FORUMSView All (14)
Ask a New Question

1

Reply
-Karuppasamy
a year ago
PDU HVDC buss Fault status
HVDC did not charge correctly in the allotted time Error code is 260100093 Gantry back down side semi conductor fuse 700V blown .So we replaced But again error code 260100093 is coming Expert’s please guide me how to solve the issue

Equipment: GE Healthcare - BrightSpeed Series

Reply

2

Replies
-toniy
3 years ago
CT brigthspeed
Please i got trouble whith tilt of the.ct brigthspeed i need some help about the tilt failureReply

1

Reply
-Mrbest
3 years ago
The table moves up but does not move down completely

GE BRIGHTSPEED SELECT CT ( 8 slice)

Error code: 260112001

It gives the error message

Interference updates are too low

Table type : 132897 sides: ox53214 states: 382000

Tilt: 0 mdeg

Elevation: 1 uml Cradle: 0 um

IMS: 340500 umh

The system is not scaning you can use the service switch on the GTCB to move the table up or down but the table will not move up or down when you press the buttons on the gantry cover.

I need help quickly patients are waiting

I have also done table characterization and it passed then I noticed that if you raise the table to the height 239 mm with the switch on the GTCB and press the button on the gantry cover the table will move up but will not go below the 239 mark. I can't even do fastcal

 

Reply
SERVICE COMPANIESView All CT Scanner Companies

FEATURES

[list] [*] Superb CT capabilities [*] Powered by VCT Technology [*] Ultra-compact size [*] Lower life-cycle costs [/list]

SERIES MODELS

GE HealthCare - BrightSpeed Edge Select

GE HealthCare - BrightSpeed Elite

GE HealthCare - BrightSpeed Elite Select

NEWSView All

BrightSpeed Elite with ASiR

A new standard in CT> GE Healthcare continues its commitment to delivering high-image quality with lowered dose with the introduction of BrightSpeed Elite with ASiR (Adaptive Statistical Iterative Reconstruction). ASiR, a new technology, delivers up to 40% reduction in radiation dose along the entire body with no compromise in image quality. With ...

FDA AlertsView All

Class 2 Recall BrightSpeed Excel/Edge/Elite Select Computed...

Failure to properly document the CTDI in the Technical Reference or User Manual. Download

Class 2 Recall: GE BrightSpeed Excel/Edge/Elite Select / Elite...

GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems. Download