Beckman Coulter - UniCel DxH 800 Coulter
by Beckman Coulter




Today’s hematology laboratories are challenged by ever-increasing pressure to improve productivity, decrease turnaround time and reduce overall costs. The UniCel DxH platform is designed to meet these challenges head-on and transform your laboratory, with unparalleled quality of results, innovative efficiency solutions and revolutionary scalability. Heralding a new era of hematology solutions, the UniCel DxH 800 delivers the ultimate in high-definition cellular analysis. This is just the beginning of the transformation journey. Partner with us to see what the future holds.
1Reply2 months ago | What is the methodology for DXH 800 for HB What is the methodology for DXH 800 for HBReply |
2Replies4 years ago | software We purchased a used DXH800 from ebay for our students in a MLT program. I am missing the software. Does anyone know where to download it from or is willing to share it? Please help!Reply |
0Replies6 years ago | Lovelle Rempel Moving Average . how to use it? Also as Beckman is having problem with body fluid QC and we have to calculate our SD , can you tell me how to format it. in SEtting. About Moving average is it a CAP requirement or not? Reply |
Height | 38 in |
Length | 31 in |
Weight | 260 lbs |
Width | 30 in |
Throughput
Up to 100 samples/hr
Aspiration Volumes
Open/Closed Vial 165 µL
Power
90–264 VAC and 48–62 HZ
DxH 800 System with System Manager and printer draws 1,185 watts
DxH 800 System with System Manager, printer and UPS draws 2,390 watts
Data Management
Computer Microsoft* Windows* XP
Database 40,000 Patient results with Graphics
Quality Assurance QC with Levy-Jennings Graph, XB/XM for Moving
Averages, Daily Check, Intelligent Quality Monitoring
Sample Transport Capabilities
Capacity 20 five-tube cassettes
Bar Codes Digital Bar Code with 2D Bar Code capability
Sample ID Up to 22 characters
Problem: In a February 11, 2011, Urgent Product Correction letter submitted by an ECRI Institute member hospital, Beckman Coulter states that the above analyzers may experience the following problems: (1) Specimen tubes may fall out of a type A cassette (product number A53551) when the analyzer inverts the cassette in the mixing station or ...
Beckman Coulter is recalling the DxH 800, DxH 600, and DxH 900 Hematology Analyzers due to sporadic erroneously elevated platelet count results without flags or system messages, meaning there is no way for the laboratory operator of the test to recognize the error.