Zoll - R Series
Follow Us
DESCRIPTION
Defibrillator
The R Series® introduced Simple, Smart, and Ready defibrillation to hospitals. Two models now make code-readiness possible for every hospital department, with advanced, cost-effective solutions for resuscitation.
FORUMSView All (12)
2Replies24 days ago | fails readiness test am obviously overlooking something here. unit had the the OK readiness check, but yet failed the self-test overnight . all tests good except ' electrode : not connected' has new paddles/one step cable. have tried the one step cable plugged into the side test port - same overnight results.... Reply |
1Reply4 years ago | Not Printing I have 2 units that will advance paper, but will not print. I check for correct paper. One I have even replaced the printer ribbon cable and the printer assembly but still nothing. Any suggestions? I hate to think its the board just for no printing.Reply |
1Reply4 years ago | Check Pads I keep getting "Check Pads" every few seconds. When I set up the 30J test, the unit will not discharge, charges, but wont discharge. Forget about higher settings, unit will not discharge at any setting. I tried another multi-function therapy cable, but no luck. Anyone seen this before? Any suggestions?Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Defibrillator Companies
FEATURES
- Uniform Operating System between the two models
- Shares common electrodes, cables, paddles, and batteries
- Only one therapy cable needed for defibrillation, cardioversion and pacing no separate 3 lead cable required.
- Real CPR Help® to coach to the proper compression rate and depth during CPR as well as See-Thru CPR® to minimize pauses during CPR compressions
- The only defibrillator that recognizes the OneStep Pediatric electrode and automatically lowers the energy setting to 50J to prevent accidental overdose of energy
- Defibrillator event documentation via internal memory, as well as a removable Compact Flash card for easy data transfer and archiving.
- Conducts an automated self-test once a day.
- Patented external pacing offers superior capture at lower mean thresholds, reducing muscle artifact and ensuring more patient comfort and better patient tolerance