Replies
-LeonardoZimolo
2 years ago
HOW TO FIX ERR. COD. PB3
Hello I ask if it has ever happened to anyone to come across the error cod. Pb3 on Sigma 3-16L laboratory centrifuge and possibly how it solved. Sigma Service advised me to replace the control board, which I did, but unfortunately the problem was not solved. Now I have been advised to try to check the value of the brake resistance which should be 330 Ohm, if the value is correct I am indicated to replace the power board which has a very important cost.Reply
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Centrifuge Companies

FEATURES

[list]      SIGMA 3-16P  [*] Universal table top centrifuge; speed range up to 14.500 rpm       [*] Various angle and swing-out rotors available       [*] Maximum capacity of 4 × 250 ml down to microtubes with different adapters in the same bucket (13190)       [*] Maintenance-free induction drive motor. A microcontroller controls speed, time, temperature and gravitational field       [*] Suitable for low speeds, e.g. for cell sedimentation, accuracy +/– 1 rpm       [*] Magnetic rotor identification prevents the rotor from overspeeding        [*] Stainless steel bowl       [*] Imbalance switch       [*] Produced according to national and international safety regulations    SIGMA 3-16PK  [*] refrigerated version of 3-16P       [*] Universal refrigerated table top centrifuge; speed range up to 15,300 rpm       [*] Possibility of precooling the rotors during standstill       [*] Refrigerant CFC-free (R 134 a)       [/list]

SERIES MODELS

Sartorius Group - 3-16P

Sartorius Group - 3-16PK

This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree