Higher-capacity laboratory centrifuge for swing-out and angle rotorsFORUMSView All (1)
2Replies2 years ago | HOW TO FIX ERR. COD. PB3 Hello I ask if it has ever happened to anyone to come across the error cod. Pb3 on Sigma 3-16L laboratory centrifuge and possibly how it solved. Sigma Service advised me to replace the control board, which I did, but unfortunately the problem was not solved. Now I have been advised to try to check the value of the brake resistance which should be 330 Ohm, if the value is correct I am indicated to replace the power board which has a very important cost.Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Centrifuge Companies
FEATURES
[list] SIGMA 3-16P [*] Universal table top centrifuge; speed range up to 14.500 rpm [*] Various angle and swing-out rotors available [*] Maximum capacity of 4 × 250 ml down to microtubes with different adapters in the same bucket (13190) [*] Maintenance-free induction drive motor. A microcontroller controls speed, time, temperature and gravitational field [*] Suitable for low speeds, e.g. for cell sedimentation, accuracy +/– 1 rpm [*] Magnetic rotor identification prevents the rotor from overspeeding [*] Stainless steel bowl [*] Imbalance switch [*] Produced according to national and international safety regulations SIGMA 3-16PK [*] refrigerated version of 3-16P [*] Universal refrigerated table top centrifuge; speed range up to 15,300 rpm [*] Possibility of precooling the rotors during standstill [*] Refrigerant CFC-free (R 134 a) [/list]