Roche - Cobas 6000
by Roche

Follow Us
DESCRIPTION
Tailor-made based on individualized workflow analysis of customers laboratory and available in 7 configurations.

The cobas 6000 analyzer series is the entry into a new family of analyzers for the clinical laboratory, the cobas modular platform. It allows tailor-made solutions for medium workload laboratories; either integrating clinical chemistry and immunochemistry testing or offering dedicated workplaces.                     

Cobas 6000 analyzer series is tailor-made based on individualized workflow analysis of customers laboratory and available in 7 configurations. Through on-site upgrades, its modular design can grow with changing demands.The test throughput of cobas 6000 analyzer series ranges from 170 up to 2170 test per hour, depending on the configuration. As a precondition for consolidation it offers up to 155 reagents in direct access.It is a highly compact and efficient analytical system that provides efficiency, convenience, safety, and flexibility.

FORUMSView All (30)
Ask a New Question

0

Replies
-osmanc
a month ago
feel free to ask

i am an engineer and working on cobas devices you can ask me anything you want, feel free to ask. if you need help just write via mail also you can call either

osman838@gmail.com

Reply

1

Reply
-sankaranarayanan.m_8147
6 years ago
Wrong result received from Cobas 6000 analyzer
Wrong result received from Cobas 6000 analyzer. For example, In the anaylyzer , wrokplace -> Data review screen, showing the result value for ALT test as 19.2, but the same samples result received in Host LIS is 319.2, Another example is in analyzer data review screen ,its showing 100 for TSH test, but the same received as 1020.0 in Host. We are getting much confused and not able to proceed further, because of this peculiar result values. Please help us to sort out this issue. Note: We are running bidirectional mode interfacing.Reply

1

Reply
-sankaranarayanan.m_8147
6 years ago
How to calculate the checksum value
Please help us and share with us the logic to calculate the checksum values to send from host to analyzer. we are struck in that process, please help us.Reply
DOCUMENTS / MANUALSView All

FEATURES

Improved efficiency

  • Cobas 6000 analyzer series can consolidate more than 95% of the routine workload of clinical chemistry and immunochemistry testing.

Greater flexibility

  • Cobas 6000 analyzer series is tailor-made based on an individualized workflow analysis of the laboratory. Through on-site upgrades of up to 3 modules in 7 different combinations, its modular design can grow with the laboratory, securing the investment in the future.

Improved safety

  • Data management capabilities via internet link, such as full traceability of work area internal processes, full system data backup, direct access to latest updates of application information for reagents, calibrators and controls and additional service options supports the evolving standard of good laboratory practice. e-services ensuring operational convenience and safety for around the clock operation.

More convenience

  • The unique concept of the Cobas c pack and the Cobas e pack with a large variety of ready-to-use reagents allows convenient reagent handling and in over 90% of all reagents calibration is required only once per lot.

SPECIFICATIONS

Height130 cm
Sample Capacity150 Samples/ 30 racks on 2 trays
Throughput600 samples/hr
Length 98 cm
Weight425 - 1,000 KG
Sample Volume1 to 35 µL
Width1850 - 4,250 cm

Additional Specifications

[list]      Sample material: Serum, Plasma, Urine, CSF      Sample dilution: 3 - 121 times, diluent > 100 µl      Noise output: < 68 dB      [/list]