DESCRIPTION
Burdick Atria 3100 12-Lead Resting ECG

The Atria 3100 ECG monitor is a standard electrocardiogram system for private practices (see the Burdick Atria 6100 for additional features). It follows on the success of our earlier model, the Burdick Atria 3000. Our technology assures an accurate report and gives ECG protocol choices, helps eliminate transcription errors, and improves workflow.

FORUMSView All (14)
Ask a New Question

7

Replies
-Vallarie
5 months ago
EKG showing abnormal on most of our patients
The last few times nearly all of our EKGs were showing abnormal. Our patients exhibited no signs or symptoms of distress. We were doing routine EKGs. We have had this machine for years. Does it need to be calibrated? If so, can we do it ourselves? Reply

3

Replies
-PRO-MED
a year ago
Performing Reliability Checks
The Atria 3100 display keeps cycling between ATRIA and Displaying Reliability Checks. What could be causing this?Reply

2

Replies
-JTBmedrepair
3 years ago
Faded printing in same areas
burdick atria 3100 getting faded printing in same area.Reply
DOCUMENTS / MANUALSView All

FEATURES

  • Stores 50 ECGs    
  • Manual or automatic 3-, 4-, 6-, and 12-lead ECG printouts    
  • Standard with wired Ethernet, wireless 802.11, and Bluetooth; complies with FAA requirements    
  • Bi-directional communication with our HeartCentrix connectivity software    
  • AccuPrint feature restricts ECG printing if lead wires are not properly attached    
  • Option to use the University of Glasgow ECG interpretation algorithm    
  • Warranty: 3-Year Parts and Labor

SPECIFICATIONS

EKG/ECG TypeRest
Height5.5 in
Length 13.1 in
Weight11 lbs
Width15 in

Additional Specifications

DATA STORAGE:    
50 records    
NETWORK CONNECTION    
Wired:    
10/100 MBPS Ethernet via RJ45    
Wireless:    
802.11g (compatible with 802.11b networks), security options including WEP (64/128 bit), WPA / WPA2 (Personal and Enterprise, including AES), EAP-TLS, PEAP/MSCHAPv2, LEAP, RADIUS servers (including Microsoft IAS, Cisco ACS and Cisco Aironet 1200), roaming across WAPs of various configurations and support for mixed 802.11 / Ethernet networks.    
POWER REQUIREMENTS    
AC operation:    
115/230 V AC +/-10%, 50/60 Hz at external power source    
Battery operation:    
14.4 V NiMH rechargeable battery pack    
Battery duration:    
300 pages continuous printing without communications enabled    
PRINTOUT    
Printout device:    
216 mm thermal dot array    
Paper dimensions:    
8.5 in x 11 in (US letter) or 210 mm x 300 mm (A4)    
Paper type:    
Thermal sensitive (Burdick Assurance or Heartline paper recommended)    
Chart speeds:    
12.5, 25, 50 mm/sec    
Gain:    
5, 10, 20 mm/mV chest or limb (may be split)    
Printout formats:    
3-, 4-, 6-, or 12-channel; additional rhythm formats    
ACQUISITION    
Lead selection:    
Meets IEC 60601-2-51    
I, II, III, aVR, aVL, aVF , V1, V2, V3, V4, V5, V6, V4R; supports Cabrera and alternate chest lead (chest lead selection V2R thru V9R + V7,V8,V9)    
Measurement (standard),    
Interpretation (optio:    
Diagnosis, measurements and reason statements based on five demo graphic criteria    
Pacemaker display capability:    
Meets or exceeds ANSI/AAMI EC11-1991    
Modes:    
Automatic, rhythm or manual    
Frequency response:    
Meets or exceeds ANSI/AAMI EC11-1991 standard    
Input impedance:    
Meets or exceeds ANSI/AAMI EC11-1991 standard    
Electrode offset tolerance:    
+/- 300 mV    
A/D Conversion:    
8,000 samples/second, 2.5 µV LSB RTI    
Artifact filter response:    
40 Hz, -3db    
Data storage resolution:    
500 samples/second, 2.5 µV LSB RTI    
SAFETY:    
Leakage current: Patient <10 µA, chassis <100 µA; Defibrillation protection: to 5000 V, 360J    
SAFETY STANDARDS:    
ANSI/AAMI EC11-1991    
UL 60601-1    
CAN/CSA C22.2 no. 0-M91    
EN 60601    
CAN/CSA C22.2 no. 601.1S1    
EN 60601-1-2:2001    
CAN/CSA C22.2 no. 601.2.25, Amend. 1    
EN 60601-2-51:2003    
CAN/USA C22.2 no. 601.1-M90    
EN 60601-1-4:2000    
DISPLAY:    
Backlit 2 x 40 character LCD    
KEYBOARD:    
Full alphanumeric keypad plus designated quick keys    
ENVIRONMENTAL    
Operating temperature:    
50°F to 104°F (10°C to 40°C)    
Operating relative humidity:    
20% to 90%, non-condensing    
Operating atmospheric pressure:    
1060 hPa to 700 hPa (-500 ft to 10,000 ft reference to sea level)    
Storage temperature:    
-4°F to 115°F (-20°C to 45°C)    
Storage relative humidity:    
10% to 90%, non-condensing    
Storage atmospheric pressure:    
1060 hPa to 190 hPa (-500 ft to 40,000 ft reference to sea level)    
WARRANTY:    
3 years with return of warranty card 