CareFusion - Alaris MedSystem III 2863
Follow Us
DESCRIPTION
The Alaris IVAC Medsystem III 2863 is an infusion pump that can be programmed to deliver at a specified rate or over a specific period of time.The Alaris IVAC Medsystem III 2863 is an infusion pump that can be programmed to deliver at a specified rate or over a specific period of time.
FORUMSView All (2)
1Reply8 years ago | Battery capacity and battery charge current I have a MSIII model 2863 with a customer complaining of short battery life. The device charge current is 180mA, which is in spec according to the manual. Batteries charged 24 hours in the device have a very low capacity. Testing the batteries in a Cadex analyzer with charge current set to 180mA, and the batteries never reach full charge. Increasing charge current to 500mA increases battery capacity to 100%. I've found this to be the case on numerous batteries, and they are all R&D. Any ideas?Reply |
4Replies8 years ago | Host system requirements for the MEDSYTEM III Field Maintenance Software? What are the host system requirements for the MED SYstem III Field Maintenance Software that is used to support the MEDSYSTEM III Infusion Pumps? Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Infusion Pump Companies
Volume to be Infused | 0.1 to 999 ml |
Height | 7.87 in |
Length | 2.1 in |
Weight | 5.1 lbs |
Flow Accuracy | 5 % |
Infusion Pump Type | General-Purpose |
KVO Range | 0.1 to 20 ml/h |
Width | 6 in |