Smiths Medical - Medfusion 2010
DESCRIPTION
The Medfusion 2010 Syringe Pump is an ambulatory syringe pump that runs on either AC current or on rechargeable internal batteries.This pump is recommended for intravenous or arterial infusions of blood, lipids, fluids, antibiotic, and any medication requiring precisely-controlled infusion rates, including epidural infusions.
FORUMSView All (9)
1Reply4 years ago | How to calibrate force sensor? How to calibrate force sensorReply |
1Reply8 years ago | Medfusion 2010 syringe pump programming We use medfusion syringe pumps to deliver boluses of medications during bypass. Our biomed tech. gave us an old model 2010 pump. The problem is the pump is programmed in a way which the ml/hr setting doesn't allow us to deliver a set volume bolus. We can set a bolus volume to be delivered but when the pump reach that volume it continue to run till syringe is empty. When I get in the program to change the settings it tkes me to screen which says Custom Program 036. The instruction manual doesn't say anything about these custom settings. Does anyone have any ideas what we can do to fix this? Thanks Allen Millar. Reply |
2Replies10 years ago | PM procedure Does anyone have a copy of the PM procedure for the 2001 & the 2010? We were supposed to have gotten rid of the last of these but 4 have shown up in a lab area & I am going to have to take care of them. Thanks. Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Infusion Pump Companies
FEATURES
[list] [*] Ambulatory syringe pumps that run on either AC current or rechargeable internal batteries [*] Accepts a broad range of syringe brands and sizes and it has automatic sensing of syringe size [/list]
Volume to be Infused | 0 to 9999 ml |
Height | 3 in |
Length | 7.5 in |
Flow Rate Range | 0.1 to 378 ml/h |
Weight | 2.5 lbs |
Flow Accuracy | 3 % excluding variations in syringe accuracies |
Infusion Pump Type | Ambulatory |
KVO Range | 0.1 to 9.9 ml/h |
Width | 4.5 in |