Zimmer - ATS 750
Follow Us
DESCRIPTION
Tourniquet
The Zimmer ATS 750 Automatic Tourniquet is a lightweight unit and very user friendly. It uses ambient air instead of using nitrogen tanks. The microprocessor controls provide accuracy and speed. The ATS 750 can be used on the countertop or with a pole mount. This unit is used.
FORUMSView All (8)
0Replies3 years ago | Black cube component in the circuit board We have a faulty tourniquet which we're trying to fix and after testing we are suspicious of a black cube that we don't really know what it is or does. We are searching all over but nothing comes up. The box has the following printed in the top: P TSD-1060 M 9932 Does anyone know something?Reply |
2Replies3 years ago | ATS failure Hi there! So, I have a trouble with my ATS 750. When I try to use it (plugged) there's no function at all -not the display, neither the pressure- but a little gasp-like sound. When doing the maintainance test, the second one -Cr14- I have no volts reading. Have anyone else had a trouble like this, or any guess to get te solution?Reply |
1Reply7 years ago | AMP FAIL What causes this?Reply |
DOCUMENTS / MANUALSView All
FEATURES
- Controls: On/Standby switch, pressure switch, time switch, cuff inflate and alarm silence switch.
- Display: Pressure - Red 14 segment light emitting diodes (LED), Time - Red 14 segment LED, AC Indicator light - Green LED.
- Cuff Pressure Range: 50-475 mmHg, 5 mmHg increments
- Pressure Accuracy: +/-5 mmHg (50-475 mmHg)
- Pressure Regulation: +/-6 mmHg of set point (10 second average under non-transient conditions without external leaks)
- Maximum Pressure: 475 mmHg cuffs
- Time Alarm Set Range: 0-240 minutes, 5-minute increments
- Timer Accuracy: 0.25% of elapsed time
- Internal Diagnosis: Program, memory, watchdog timer, transducer calibration, improper valve actuation
Additional Specifications
- Height: 8 in (20.3 cm),/li>
- width: 7 in (17.8 cm)
- Depth: 7.9 in (20.1 cm)
- Weight: 7.7lbs (3.5 kg)