Pronk Technologies - OX-1 OxSim
Follow Us
DESCRIPTION
Miniaturized Optical SpO2 Pulse Oximeter TesterTest a whole system, including the sensor, with the OxSim® Optical SpO2 Pulse Oximeter Tester. Slips into your shirt pocket and operates easily with just two buttons. Access 4 saturation values, 3 heart rates, and 2 perfusion levels. Utilizing optical technology, the OxSim® SpO2 simulator tests your SpO2 monitoring systems thoroughly, quickly, and accurately.
FORUMSView All (1)
0Replies5 years ago | BMET Pack PRO for On-the-Go Biomeds Pronk Technologies introduces BMET Pack PRO: a compact and versatile package that makes it easier than ever to travel with your test equipment between hospitals and clinics. Reply |
DOCUMENTS / MANUALSView All
VIDEOS
BMET Pack PRO for On-the-Go Biomeds
FEATURES
- Simple operation: Just two buttons
- Convenient: No adapter cables to purchase or carry
- Strong: Tested for ruggedness. 65 drops from a height of 3.5 feet (1.1 m) to a hard surface with no failures
- Efficient: 8 hours of continuous simulation on a single 'AA'
battery. Compatible with the Battery Boost run OxSim & SimCube NIBP Simulator simultaneously.
Weight | 5 oz |
Analyzer/Simulator Type | Pulse Oximeter |
Additional Specifications
OxSim® Optical SpO2 Pulse Oximeter Tester Features:
- SpO2 Simulations:
- Mode 1:
- Saturation: 85%
- Heart Rate: 80 bpm
- Perfusion Index: approx. 2.0
- Mode 2:
- Saturation: 95%
- Heart Rate: 40 bpm
- Perfusion Index: approx. 2.0
- Mode 3:
- Saturation: 98%
- Heart Rate: 80 bpm
- Perfusion Index: approx. 2.0
- Mode 4:
- Saturation: 98%
- Heart Rate: 140 bpm
- Perfusion Index: approx. 2.0
- Mode 5:
- Saturation: 99%
- Heart Rate: 80 bpm
- Perfusion Index: approx. 0.2
- Mode 1:
- Size:
- 2-1/2″ x 2-1/2″ x 1-1/4″ (6.4 x 6.4 x 3.2 cm), excluding finger section
- Weight:
- Only 5 oz. (0.14 kg, excluding battery)
- Simple operation:
- Just two buttons
- Convenient:
- No adapter cables to purchase or carry
- Economical:
- ‘Finger’ is fixed and permanent. No parts to carry or replace annually
- Rugged:
- Polycarbonate housing. Drop tested 65 time from a height of 3 feet (1m) to a hard surface with no failures!
- Power:
- 6VDC Adapter or AA battery. Compatible with the Battery Boost — run OxSim & SimCube NIBP Simulator simultaneously.
- Battery Life:
- Minimum 8 hours. 60 days in sleep mode (unit sleeps after 10 min. with no probe connection)
- Warranty:
- Four-Year Limited Warranty
Package Contents:
- (1) OxSim® Optical SpO2 Pulse Oximeter Tester
- (1) Power Supply (6VDC)
- (1) AA battery
- (1) Certificate of Calibration
- (1) Carrying Case