The compact AK 96 machine enables dialysis providers to consistently deliver the highest level of quality and safety in hemodialysis (HD) treatment with improved cost-efficiency. Innovative features such as the Diascan system and a completely new user interface offer a simpler way to consistently deliver flexible high quality, treatments for patients.
FORUMSView All (40)
1Reply9 months ago | regarding the availability need to have the suppliers for the product Equipment: Gambro(Sweden) Flow Pump Unit PN:K18998001 for GAMBRO AK 96(New,Original) Reply |
1Reply2 years ago | Please help C CFF 018 005 BUBBLE CHAMBER (DEAIRATING CHAMBER) ERROR Unconditional error, blood pump available General description of conditions for occurring: - AIVA and REVA opened. Suction pump active and fluid tubes connected to safety bypass. Low level in bubble chamber is not achieved within 90 sec. The status from level sensor indicates middle level in chamber. Technical description of conditions for occurring: - PRL(P_BubbleTrapHigh)=FALSE Some possible explanations why the error has occurred: - The AIVA or REVA could not be opened - Broken float in bubble chamber - Broken protective driver board (IC28)Reply |
1Reply2 years ago | C CFF 018 005 BUBBLE CHAMBER (DEAIRATING CHAMBER) ERROR Please help what to do C CFF 018 005 BUBBLE CHAMBER (DEAIRATING CHAMBER) ERROR Unconditional error, blood pump available General description of conditions for occurring: - AIVA and REVA opened. Suction pump active and fluid tubes connected to safety bypass. Low level in bubble chamber is not achieved within 90 sec. The status from level sensor indicates middle level in chamber. Technical description of conditions for occurring: - PRL(P_BubbleTrapHigh)=FALSE Some possible explanations why the error has occurred: - The AIVA or REVA could not be opened - Broken float in bubble chamber - Broken protective driver board (IC28)Reply |
DOCUMENTS / MANUALSView All
FEATURES
[list] [*] Diascan - Controlled dialsysis dose delivery [*] Blood pressure monitor - Integrated real-time blood pressure monitoring via patient cuff [*] U9000 dialysis fluid ultrafilter - Ultrapure dialysis fluid for increased biocompatibility [*] Arterial blood line clamp and variable dialysate fluid flow - For added economy, reliability and flexibility [*] Battery back-up - For safe treatments in allenvironments [/list]