Arizant - Bair Hugger Model 775
by Arizant

Follow Us
DESCRIPTION

Patient Warming Unit 

Arizant's latest advancement in forced-air warming - the Bair Hugger Model 775 temperature management unit - is the future gold standard in patient warming.

FORUMSView All (10)
Ask a New Question

1

Reply
-DPerkins28
10 months ago
Heater Replacement

Anyone have procedure on how to change the heater in this unit?

Reply

0

Replies
-jcarpen
10 months ago
FC-001

Currently troubleshooting Bair Hugger 775. Getting fault code 001. Device displays over temp. Switched hose and ran it on another bair hugger and it did not display an error. I have eliminated the hose being the issue.  Any tips or suggetions? Thank you!

Equipment: Bair Hugger 775

Reply

1

Reply
-Know
a year ago
Fc 009,Fc 008 Fc010

White residue on control board? Will this cause a fault? Very high resistance between white and yellow at connector J4

Equipment: Arizant - Bair Hugger Model 775

Reply
DOCUMENTS / MANUALSView All

FEATURES

  • Adjustable airflow, precise temperature delivery and ease of use from a single, quiet warming unit
  • Two airflow settings and hose-end temperature sensing ensure precise temperature delivery and compatibility with all Bair Hugger blanket styles - pediatric to adult
  • Three temperature sensors and adjustable airflow provide safe, quick temperature response and greater heat transfer
  • Tracks and reports forced-air warming therapy duration, over-temperature information, calibration, built-in hour meter and fault codes all via the front panel - no need to open the unit
  • Compatible with Bair Hugger 241® blood/fluid warming set
  •  High-efficiency 0.2 μm air filter
  • Safe, quick temperature response
  • Adjustable airflow, precise temperature delivery, and ease of use from a single, quiet warming unit
  • Compatible with all Bair Hugger Blanket Styles - pediatric to adult

Additional Specifications

  • Length: 13" (33 cm).
  • Width: 14" (36 cm).
  • Depth: 13" (33 cm).
  • Net Weight: 16 lb (7.3 kg).
  • Brand: 3M Bair Hugger Therapy.
  • Clinical Area: OR, PACU, Pre-Op.
  • Device Rating: 110 - 120 VAC, 50/60 Hz, 11.7 Amperes.
  • Filter Type: High-efficiency 0.2 μm air filter.
  • Leakage Current: Meets UL 60601-1 and IEC 60601-1 requirements.