Philips - Mx8000 IDT 16
by Philips

Follow Us
DESCRIPTION

CT Scanner

The sophisticated and refined performance of the Mx8000 IDT 16 is now within your control to apply ingenuity to difficult diagnostic challenges. The Mx8000 with Infinite Detector Technology is optimized for isotropic detail and high image quality with low dose protocols.

FORUMSView All (5)
Ask a New Question

1

Reply
-Matroni
Sat Sep 12 2020
Detectors problem

Please help, I have problem with ring artefacts on my Philips mx8000. I done Bad detector test but I cant solve this .Can somebody give me some advice, I can send images what I have done.

Reply

1

Reply
-ESTDK
Thu Jan 09 2020
What is the liquid inside CT phantom?
We are trying to dispose of a PHILIPS MX8000 IDT BRILLANCE CT PHANTOM, 455012201471, but cannot get information out of Philips Medical as to the liquid inside the phantom. It is thicker than water so there must be a SDS or other specification on it for proper disposal. Does anyone have documentation on what kind of liquid is inside the phantom? Philips customer service support so far has been unable to help with the needed information.Reply

1

Reply
-Unicgold
Sun May 15 2016
Ver 4.2.0/2.2

when ever the gantry is attempted to be control from the 3.2.0 on the host it corrupt the software and the gantry does not come ready with ver3.2.0, Please we need help

Reply

FEATURES

  • Asymmetrix™-Philips Medical System’s Patented variable wide area detector providing optimal dose efficiency with 24mm total coverage 
  • Excellerator 800 Mbt/second Ultrahigh speed multislice data acquisition system
  • 16 slice acquisition mode – 16 x 0.75mm, 16 x 1.5mm, 8 x 3mm scanning 
  • 0.5 second scan time for 360°degree scan
  • Up to 6 images/second reconstruction time, 512 x 512 image
  • On-board 60kW, high frequency, high-voltage generator
  • 6.5 MHU high power x-ray tube
  • Dynamic Focus System (DFS) doubles data density providing up to 24 lP/cm ultra-high Spatial resolution, in axial and spiral scanning 
  • DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation) 
  • Gantry and table controls located on both sides of the gantry and on the operators console 
  • Large 50 cm field-of-view inside a wide, flared 70 cm aperture 
  • Multiple-Slice, volumetric spiral scan: up to 97.5s continuous, multiple, bi-directional acquisition 
  • UltraImage 
  • Bolus Pro Ultra, Spiral Auto Start, Evolving Image, Ultra High resolution matrices
  • Patient couch with 1570mm scannable range 
  • Real-time image processing: Zoom & Pan, cine, Multiformat and Image Graphics 
  • 80cm spiral coverage in one 13 sec. breath-hold with 1.5mm slice thickness
  • Windows XP Dell Precision host computer     
  • DICOM 3.0 complaint image format, archive and networking

SPECIFICATIONS

Type (configuration)Multislice
Slices16
Weight2100 kg
Gantry (Max. Clearance) 70 cm
Image Display21 in crt
This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree