The MAC 5500 system offers the sophistication required for advanced ECG applications, while its ease of use extends this level of performance to the broadest range of users possible. And, it's part of the complete GE suite of networked, non-invasive testing solutions designed to maximize efficiency and productivity in hospitals, clinics, office-based practices, and research institutes.
2RepliesWed Oct 29 2025 | Bar code scanner not working Hi all, When pressing the F1 key to select "patient data" on the MAC 5500, there is this quick flash to scan the patient visit/ID number to add them to the ekg cart, then it switches to manual entry right away. Kept pressing that F1 button to use the scan function, but it will switch to manual entry. What's causing the issue? Software upgrade? Internal board malfunctioning? Settings? I've double checked the settings to another working ekg and all settings matched. Problem still there. Tried hard reboot of device. Nothing. Replaced the scanner and cable to a known working one. Same issue. Can't get in contact with GE tech support. Anyone has had this problem? Reply |
2RepliesTue Jul 08 2025 | Printing maybe Error when printeing why Reply |
0RepliesFri Jul 04 2025 | Where do you go to find steps to troubleshoot it verygoodo ha nevermind Reply |