4

Answers
-Crutch
2 months ago
LL and RL leads
Got a machine, on a patient simulator, the unit displays LL disconnected.. Changed the LL cable, LL disconnected. Changed the banana clip end, LL disconnected. Changed the CAM module, LL disconnected. Changed the CAM Cable, LL disconnected. Put the RL on the LL, unit functions normally. Thoughts?Reply

1

Answer
-JV
4 months ago
need password for Mac 5500
I replaced the print head yesterday. I need password to input resistance value of new printhead. I'm guessing there is a standard password to enter the SETUP page.Thanks for your helpReply

7

Answers
-RJBiomed
4 months ago
Display Blank
Hi Folks, I've got a MAC 5500 with a blank display. I shone a flashlight across the screen and I could make out text on the screen. I swapped the display with a new one out of the box but the blank display issue continued. I tried the new display on another machine and there were no issues. Could the main board be the problem? It's usually a cable issue. Thanks in advance.Reply
DOCUMENTS / MANUALS
PREVENTIVE MAINTENANCEView All
Be the first to submit a preventive maintenance post about this equipment

FEATURES

          
  1. Ethernet and wireless networking option saves time, reduces errors and maximizes charge capture      
  2. Hook-up advisor minimizes editing time and repeat ECG acquisitions      
  3. Single button operation for printing, storage and transmission      
  4. Bar code scanner helps reduce clerical errors      
  5. 12SL ECG Analysis program with Gender-Specific software      
  6. 15-Lead ECG Analysis      
  7. Risk stratification tools such as ACI-TIPI, Late Potential P-Wave Signal Averaging      
  8. Optional stress testing capabilities      
          

SPECIFICATIONS

EKG/ECG TypeRest
Height3.37 in
Length 13.8 in
Weight15 lbs
Width15 in
NEWSView All
!{article.title}
GE Login Tips from Conquest Imaging

GE login tips from Conquest Imaging

!{article.title}
GE Healthcare Launches the Most Sensitive Clinical PET/CT Scanner in...

GE Healthcare is proud to announce the new Discovery† PET/CT 610 as the first double-digit sensitivity PET/CT scanner* with an absolute sensitivity of 10cps/kBq1, the highest level found on any scanner commercially available.

Advertisement

Ask a Question

Ask anything about
GE Healthcare - Marquette MAC 5500
Advertisement

For questions about using the MedWrench web site:

(866) 989-7057TOLL FREE  •  (770) 692-8371INTERNATIONAL

support@medwrench.com

MedWrench is in association with:

TechNationMedical DealerOR TodayMD Expo


Testimonials  •  Advertising  •  Privacy Policy  •  Terms and Conditions

© 2017 MedWrench, LLC. All Rights Reserved

Advertisement