Bovie - Aaron 2250
by Bovie

Follow Us
DESCRIPTION
Digital Electrosurgical Generator

The Aaron 2250 is a multipurpose electrosurgical generator for use in the modern operating room and surgi-center. It features both monopolar and bipolar functions to satisfy all of your surgical demands with safety, flexibility, reliability, and convenience.

FORUMSView All (4)
Ask a New Question

0

Replies
-rymem2669
Fri May 30 2025
Manual servicio
Error 9 pienso que es el oxilador

Equipment: Bovi AAron 2100

Translation: Error 9 I think it's the oxylator

Equipment: Bovi AAron 2100

Reply

4

Replies
-Nick Philbrick, CBET
Thu Jun 16 2016
Bovie 2250 E-4 Error Message DC Power Supply Help

Good Morning,

We have a Bovie 2250 that gives us a random E-4 message during use. The service manual says that the E4 error is a DC power error and has no additional information. I replaced the small low-voltage DC power supply board as it was only $70 and ensured that all of my low-voltage DC power levels were adjusted appropriately and the error seemed to subside. After the ESU was back into service, the team supposively had the error return. Does anyone have any experience with this or a more in depth error guide?

Thanks, 

Nick Philbrick, CBET

Facilities and Medical Equipment Manager

Veterinary Specialty Hospital

6405 Tryon Rd, Ste. 100

Cary, NC 27518

Direct: 919-600-6605

Hospital: 919-233-4911

 

Reply

0

Replies
-@hmm
Fri Sep 26 2014
bovie 2250

good moring how to eliminate the error code E2 in bovie aaron 2250

Reply
DOCUMENTS / MANUALSView All

FEATURES

[list]        [*] Large illuminated digital displays        [*] Five Output Modes:        Cut (200 Watts)        Blend (200 Watts in 9 steps)         Coagulation (Pinpoint 120 Watts)         Fulguration (Spray 80 Watts)         Bipolar (80 Watts)        [*] Self-test circuits        [*] Audible tones         [*] Discreet outputs        (FCFS™ - First Come First Served)         [*] Isolated circuitry         [*] BovieNEM™        (Neutral Electrode Monitoring)        [*] Digital Error Detection        [/list]

SPECIFICATIONS

Height6 in
Length 16.3 in
Weight19 lbs
OutPut Frequency490 kHz ± 5
Output Power200 W Cut/Blend
Width12.3 in

Additional Specifications

Output Power        Cut/Blend: 200 watts/200watts 10 Steps         Coagulation (Pinpoint): 120 watts         Fulguration (Spray): 80 watts         Bipolar: 80 watts                Line Voltage-100-240 VAC ± 10%                   Line Frequency-50-60 Hz