Bovie - Aaron 2250
DESCRIPTION
Digital Electrosurgical GeneratorThe Aaron 2250 is a multipurpose electrosurgical generator for use in the modern operating room and surgi-center. It features both monopolar and bipolar functions to satisfy all of your surgical demands with safety, flexibility, reliability, and convenience.
FORUMSView All (3)
4Replies8 years ago | Bovie 2250 E-4 Error Message DC Power Supply Help Good Morning, We have a Bovie 2250 that gives us a random E-4 message during use. The service manual says that the E4 error is a DC power error and has no additional information. I replaced the small low-voltage DC power supply board as it was only $70 and ensured that all of my low-voltage DC power levels were adjusted appropriately and the error seemed to subside. After the ESU was back into service, the team supposively had the error return. Does anyone have any experience with this or a more in depth error guide? Thanks, Nick Philbrick, CBET Facilities and Medical Equipment Manager Veterinary Specialty Hospital 6405 Tryon Rd, Ste. 100 Cary, NC 27518 Direct: 919-600-6605 Hospital: 919-233-4911 Reply |
0Replies10 years ago | bovie 2250 good moring how to eliminate the error code E2 in bovie aaron 2250 Reply |
1Reply11 years ago | Aaron 2250 Bovie Anyone have a service manual or output check out procedure? Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Electrosurgical Unit (ESU) Companies
FEATURES
[list] [*] Large illuminated digital displays [*] Five Output Modes: Cut (200 Watts) Blend (200 Watts in 9 steps) Coagulation (Pinpoint 120 Watts) Fulguration (Spray 80 Watts) Bipolar (80 Watts) [*] Self-test circuits [*] Audible tones [*] Discreet outputs (FCFS™ - First Come First Served) [*] Isolated circuitry [*] BovieNEM™ (Neutral Electrode Monitoring) [*] Digital Error Detection [/list]
Height | 6 in |
Length | 16.3 in |
Weight | 19 lbs |
OutPut Frequency | 490 kHz ± 5 |
Output Power | 200 W Cut/Blend |
Width | 12.3 in |
Additional Specifications
Output Power Cut/Blend: 200 watts/200watts 10 Steps Coagulation (Pinpoint): 120 watts Fulguration (Spray): 80 watts Bipolar: 80 watts Line Voltage-100-240 VAC ± 10% Line Frequency-50-60 Hz