Terumo - Sarns 8000
Follow Us
DESCRIPTION
Heart-Lung Bypass Unit
The Sarns Modular Perfusion System 8000 offers uncompromised reliability and safety. Terumo’s continued engineering efforts for the Sarns System 8000 testify to its support of perfusionists. This trusted heart-lung machine has been used in millions of cardiac surgery cases worldwide
FORUMSView All (6)
1Reply2 years ago | Service Mannaul Sevice Mannaul Required Reply |
1Reply3 years ago | Service Manuals I need operating & service manual of sarns 8000 heart lung machine. .. Thanks in advanceReply |
1Reply3 years ago | Need reservoir holder I am looking for a reservior holder for our Sarns 8000 bypass pumpReply |
DOCUMENTS / MANUALSView All
FEATURES
- Modular design
- Four or five pump base
- Roller or centrifugal pumps
- Arterial and cardioplegia monitors for temperature, pressures and timers
- Battery module
- Electronic venous occluder
- Pulse mode
- Communications module for data acquisition
Additional Specifications
- Leakage Current: <100 microamperes for 115 V and 100 V bases (fully loaded);<200 microamperes for 220-240 V bases (fully loaded).
- The maximum leakage current for the base without pumps is 45 microamperes for all input voltages
- Ground resistance: 0.1 ohms maximum for all bases
- Emergency power: 8000 system battery (two 12 volt, sealed gel cell batteries) will power the arterial pump (operating in continuous mode), the monitors, occluder and lamp for 40 minutes; when the arterial pump is operated in pulsatile mode or the cardioplegia pump is also operating on battery power, the battery time available will diminish to 25 minutes
- The base is tested for high potential: 1200V~ for 100-115 VAC units; 1500V~ for 220-240 VAC units