DESCRIPTION

MRI System

The Philips Infinion 1.5T was designed with patient care quality and high efficiency in mind. It is the shortest and most open 1.5T MRI scanner available, as well as one of the most affordable.The Philips Infinion 1.5T boasts an ultra short 1.4m long magnet, making it ideal for the most challenging patients. The Philips Infinion 1.5T achieves all this without compromising image quality and clinical performance. The Philips Infinion features the first dockable patient couch that combines a Synergy Integrated Spine Array Coil. The Philips 1.5T Infinion PD2000 is one of the shortest and most open scanners available, providing superior patient comfort and exceptional aperture.

FORUMSView All (1)
Ask a New Question

3

Replies
-Wilver
9 years ago
Horizontal motor change

I just change the horizontal motor in the couch of the system. All the movements of the couch are working correctly and the trolley reaches its home position without problem; but the system is not reconizing the motor. It won´t let me do scan becuase "the initialization stopped because an essential component is missing or has a fatal error" concerning to the ANIMATIC (THAT IS THE HORIZONTAL MOTOR)

The motor that I replace is not the exact same model, the one that is damage is model: SM2330SQ and the new one is SM2320SQ

My question is if I have to change something in the system that tells that the motor is diferent model?

Do I have to do something else than just change the motor?

Reply