Carestream Dryview 5700 Laser ImagerThe DryView 5700 Laser Imager* offers simplicity and affordability for healthcare facilities of all sizes. This innovative tabletop medical imaging system offers a modest acquisition price, which is half of the affordability equation. It also delivers consistently low operating costs. This combination creates a complete picture of true affordability. It’s ideal for computed radiography (CR), digital radiography (DR), computed tomography (CT), PACS Viewer, and magnetic resonance (MR) applications.
FORUMSView All (18)
0Repliesa year ago | optics calibration failed issue on dryview 5700 printer I got an issue with optics calibration failed on dryview 5700 printer with code 607. Does anyone have any idea about this, how to solve it, please suggest me.Reply |
6Repliesa year ago | Image unsent Issue I got an Issue with Printer DV5700. The issue was image unsent. Does anyone have an idea, how to solve it???Reply |
0Replies2 years ago | Impression film une partie en noir Bonjour madame monsieur Jai la machine Dryview 5700. Au début quand je lançais limpression . Le filme sortons propre et nickel. Actuellement, cest toujours propre sauf que une partie du filme simprime avec une zone sombre. Avez-vous déjà rencontré ce problème ?Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Medical Printers/Imagers Companies
FEATURES
[list] Laser image quality Every image on every film size is printed using true laser technology. Automatic Image Quality Control (AIQC) uses a built-in densitometer to maintain film-to-film consistency. Broad printing capabilities Print from CR, DR, CT, MR, PACS Viewers and other grayscale medical film applications. CR & DR compatibility Fully compatible with the CARESTREAM CR and DR Systems, the DRYVIEW 5700 Laser Imager is ideal for dispersed or remote printing applications from computed radiography or digital radiography systems. Ease of operation No daily or weekly preventive maintenance is required on print heads or film transport rollers to optimize image quality. A simplified user interface, room light film loading, and easy film loading further facilitate use. True dry laser imaging We created the new DryView 5700 Laser Imager for healthcare professionals worldwide who thought dry technology was too expensive or too complicated for their operations. In addition to an affordable purchase price and low operating costs, the DryView 5700 Laser Imager offers the simplicity and performance you need to improve productivity and patient care. More affordable by design Because DryView laser imagers do not use thermal print heads, there is no need for daily or weekly maintenance. And with our dry technology, there’s no need to deal with “wet” chemistry and the related environmental issues. All of this contributes to the printing system’s low cost of ownership. Automatic Image Quality Control (AIQC) Our unique AIQC technology uses a built-in densitometer to ensure films are printed according to your imaging standards. AIQC detects image quality and makes any necessary adjustments. This unique feature, combined with the proven advantages of DryView technology, ensure excellent image quality for a wide range of imaging applications. Compact design, quick installation An innovative, ergonomic design and a compact 26 x 24 in. (66 x 61 cm) footprint make the DryView 5700 Laser Imager a convenient addition — even in small workspaces. A simplified installation process ensures the printing system is easily and quickly installed. Minimal training, easy operation The DryView 5700 Laser Imager minimizes total cost of ownership through simplified operation. An intuitive user interface also speeds operation and eliminates the need for training [/list]
Height | 19 in |
Length | 26 in |
Weight | 120 lbs |
Width | 24 in |
Additional Specifications
Technology: Photothermographic (dry laser) DryView Laser Imager Quality : True laser technology; 325 pixels per inch; 78 micron laser spot spacing Throughput:Up to 45 - 85 films per hour (film size dependent). Smaller film sizes print faster Choice of Film Sizes: [*] 14 x 17 in. (35 x 43 cm) [*] 11 x 14 in. (28 x 35 cm) [*] 10 x 12 in. (25 x 30 cm) [*] 8 x 10 in. (20 x 25 cm) Network Connectivity:Integrated DICOM interface supports printing from DICOM print modalities ; Connect non-DICOM modalities using PACS Link Medical Image Managers Network connection : 10/100/1000 BaseT Ethernet connection to imager (supports jumbo frame) Power: 180–264 VAC; 50/60 Hz; maximum 4.5 amps Temperature: 59 to 91°F (15 to 33°C) Humidity: 20 to 80% RH, non-condensing Magnetic field: < 50 Gauss Altitude: –100 to 10,000 ft (–30.5 to 3,048m)