Getinge - Castle 3633
by Getinge

Follow Us
DESCRIPTION
Getinge Castle 3633 Steam Sterilizer

Getinge Castle 3633 Steam Sterilizer

FORUMSView All (1)
Ask a New Question

5

Replies
-CJaime
9 years ago
Getinge 3633 pressure in chamber

Hello, I have an old 3633 and we are getting a chamber in pressure fail once cycle completes. Once it exhausts, then it lets air in then it climbs in pressure to about 5psi when it should be zero. any help would be greatly appreciated.

Btw, parts replaced: chamber drain valve, slow exhuast rebuilt, trap replaced. all check valves replaced. Water valve rebuilt.

Reply