Lorad Mammography MIII
1ReplyWed Mar 10 2021 | service manual how can I calibrate this mammography system lorad 3-000A? Reply |
1ReplyTue Sep 10 2019 | HIGH VOLTAGE PANEL FAILURE Hello dears. I have a flaw in a Lorad III who could help me. The device presents a constant alarm at the initialization start that indicates high voltage on the panel. And verify the operation of the isolation transformer and I have voltages in the microprocessor and memory control cards (control). However, there is a short in a fuse called F2 of 5 Amp. Someone could help me in this fault please. willmary1004@gmail.comReply |
0RepliesThu Jun 02 2016 | Got "FILAMENT CURRENT FAIL" during exposure. Hi dear colleagues. I am working on a Lorad M-III Mammography system and got the "FILAMENT CURRENT FAIL" message during an exposure release using small focal spot. I read 0.566 Vdc at R42 on TUBE CONTROL BOARD. According to the service manual, it should be 0.54 Vdc. Could it be the problem? How can I adjust that voltage? Any help will be highly appreciated. Best regards. My e-mail: vicmaal@yahoo.com.mx Reply |