1

Reply
-enahavandi
Wed Mar 10 2021
service manual
how can I calibrate this mammography system lorad 3-000A? Reply

1

Reply
-willcol
Tue Sep 10 2019
HIGH VOLTAGE PANEL FAILURE
Hello dears. I have a flaw in a Lorad III who could help me. The device presents a constant alarm at the initialization start that indicates high voltage on the panel. And verify the operation of the isolation transformer and I have voltages in the microprocessor and memory control cards (control). However, there is a short in a fuse called F2 of 5 Amp. Someone could help me in this fault please. willmary1004@gmail.comReply

0

Replies
-VicMayal
Thu Jun 02 2016
Got "FILAMENT CURRENT FAIL" during exposure.

Hi dear colleagues.

I am working on a Lorad M-III Mammography system and got the "FILAMENT CURRENT FAIL" message during an exposure release using small focal spot. I read 0.566 Vdc at R42 on TUBE CONTROL BOARD. According to the service manual, it should be 0.54 Vdc. Could it be the problem? How can I adjust that voltage? Any help will be highly appreciated.

Best regards.

My e-mail: vicmaal@yahoo.com.mx