Hillrom - Advance 1105
by Hillrom


Perfect for hospitals, surgery centers and teaching facilities.
1ReplyThu Sep 07 2023 | moving transort how to fold, fit through door? Equipment: Hillrom - Advance 1105 Reply |
1ReplyTue Mar 26 2019 | Head will not raise When I raise the head of the bed from the flat position, it raises approximately 2 inches and stops. It automatically returns to the flat position as soon as the raise button is released.Reply |
1ReplySun Aug 12 2018 | Head motor gear reducer failure. The gear reducer in the head motor assembly prematurely fails. The weight of the head assembly with or without a patient seems like more than the motor design can tolerate. I have discussed this with the motor manufacturer (U.S. Motor). How can I easily test the limit switch assembly in or out of the bed to verify operation. Are there proximity sensors along the switch strip to monitor position or just end travel limits? I suspect gear damage begins when screw over travels the up limit. Thanks for any related experience help.Reply |
[list] [*] 4-Motor bed [*] Hi-Lo travel time: 21 seconds [*] Bed frame clearance w/5″ casters: 7.5" [*] Central brake and steer [*] Automatic night light [*] Instant CPR release [/list]
| Height | 17 in to 30 inches |
| Length | 91 in |
| Width | 36 in |
Sleep Surface Dimensions: 36″x 82″