Xavant Technology - STIMPOD NMS410
by Xavant Technology


Auto sensing technology provides a solution for simultaneous nerve mapping and nerve location. The Stimpod monitors whether the mapping probe or needle touches the patient and adjusts the current range accordingly, ensuring quick and precise nerve location. The Stimpod Nerve Mapping Probe was designed to enable transcutaneous nerve mapping at higher currents (to a maximum of 20mA), whilst the tip offers a contact surface small enough to ensure effective discrimination. Deeper peripheral nerves require higher current to be recruited by transcutaneous stimulation. The Stimpod sensing technology and Nerve Mapping Probe designs facilitate larger currents, thus increasing the percentage of peripheral nerves that can be targeted. The STIMPOD displays real time graphical representation of the actual waveform delivered. The displayed waveform serves as a quick reference for excessive impedance (> 20kOhm) in the circuit. The waveform will indicate if the pulse is delivered according to the settings. If the waveform is not square, this will immediately indicate excessive impedance in the circuit and all elements in the circuit i.e. ecg electrodes and/or skin condition, needs to be re-assessed before nerve location can be successfully completed.
[list] [*] Nerve Mapping Probe [*] Charge Transfer Waveform [*] Proximity Indicator [*] Non-Linear Current Adjustment [/list]
| Height | 145 mm |
| Length | 90 mm |
| Width | 30 mm |
[list] [*] Current Range - Nerve Locating: 0.0 - 5.0mA - Nerve Mapping: 0 - 20mA [*] Load Impedance - Nerve Locating: 0 - 20kΩ (100V) - Nerve Mapping: 0 - 20kΩ (400V) [*] Stimulating Frequency - 1Hz, 2Hz, 5Hz [/list]