Hi there, my machine failed to start because of the hard disk, am looking for application software for the machine so that I replace a new hard disk. Thanks

Equipment: GE Healthcare - Logiq 400

RE: GE HealthCare - Logiq 400
Added  Feb 8 2021
0 Replies

logic 400 GE ultrasound scanner 1. the machine is not switching on the screen and keyboard despite all fuses being intact. what could be the problem? 2. if its a faulty power supply where do i get one?

RE: GE HealthCare - Logiq 400
6 Replies

Hello! Sir When I power on the machine only keyboard back light is on for 1 sec then no response from the machine, same thing will happen when power off the equipment. Thanks for your response & considering this thread. Regard Zohaib

RE: GE HealthCare - Logiq 400
Updated  Aug 30 2019
1 Reply

Hello everybody,

I'm looking for help with the LOGIQ 400, Model 2265944.

Unfortunately I do not receive any support for this device in my country (Germany).

The device remains at power-up with the message at the bottom "CPU MEMORY TEST: SUCCESS", no further reactions can be detected.

In some forums I could read that this cause can be due to a defective timekeeping RAM. I got a new DS1643-150 and replaced it.

However, the system stops at boot up as with the old RAM.

Now I read in the data sheet of Timekeeping RAM that this must be programmed before its use.

So my question: How does the programming of the timekeeping RAM? Can someone help me with a guide or write where I find this?

My e-mail: m.prescher@online.de

Thanks for your time

RE: GE HealthCare - Logiq 400
Added  Jul 26 2018
0 Replies

I have Ge Logiq 400 version 2290505 that need software If anyone can send me image of the software My email fayezsami1968@gmail.com Thank you

RE: GE HealthCare - Logiq 400
11 Replies

Hello I have a LOGIQ 400 MD Version 3.30 By 2136319-4. I have replaced the Timekeeper Battery and I setted datetime, id and video values without problem. Now the systems boots but do not recognize the proves which before were working. During boot sequence appears this message on the monitor display: SYS ERROR: BOARD NON EXISTENT. BOARD: DCON And this message via hyperterminal connected with service cable: VTSH> SCNI:rt_cancdlyresume error: 11007 SCNI: USC [u_read] Poling Timeout [ffff] +++ ATS0=0 I also checked these details: * I have checked the boards installed on the system via board configuration display appears all correct except DCON or CONN(1). * I have checked wires between DCON: TRDS ASSY - HBFR (1) ASSY - HBFR (2) ASSY . Perhaps were disconected during Timekeeper fix but now appears well connected. Any ideas? Maybe I need to make some adjustments via hyperterminal to make DCON work? Maybe a problem with a Mother IC? Thanks for your time. Regards from Spain.

RE: GE HealthCare - Logiq 400
2 Replies

can anyone help me to reset bbram in logic400 after change time keeper ram battery

RE: GE HealthCare - Logiq 400
Updated  Mar 29 2017
5 Replies

I installed new IC timekeeper, but after turning on the unit, it stopped in second bar in boot up process. It seems to need input ID,date and time by hyper terminal to system,I need information about command and anything to connect laptop to unit.

RE: GE HealthCare - Logiq 400
Updated  Oct 26 2015
1 Reply

  have a problem of error code 20166  pliz assist me

      george finley

     email gmaingi559@gmail.com