GE HealthCare - Panda iRES
by GE HealthCare

GE HealthCare Panda iRES Forums

Updated  Jan 19 2024
4 Replies

Hi all,

I have the panda resus hooked up to a liner and canister. When I turn the suction on, I'm expecting a vacuum of 50 mmHg but only measure ~20 mmHg. I also try increasing the expected vacuum but it still only measures ~20 mmHg. The vacuum does increase when the liner is manually pushed down when the suction is on, but manual intervention shouldn't be necessary.  

The resus passed all checkout tests, I tried calibrating it anyways, and it passed all checkout tests again. 


I checked for cracks or leaks in the liner and canister and didn’t notice any. Tried replacing the liner and canister with another set and that didn't resolve my problem. I also tried both the original set and the replacement set on another vacuum and they both worked as expected without manual intervention.

 

Any recommendations for what to try next? Tests, replacement parts, or something else?

RE: GE HealthCare - Panda iRES
Updated  Dec 20 2023
7 Replies

Hello all,

The screen on my Panda will not light up. 

I can see the screen when I shine a flashlight on it.

Is it just the backlight or is there something else involved?

RE: GE HealthCare - Panda iRES
Updated  Nov 30 2023
1 Reply

Troubleshooting: just replaced PCA board, still getting same error: sys failure 09

Equipment: GE HealthCare - Panda iRES

RE: GE HealthCare - Panda iRES
0 Replies

I have a panda that the procedure lights intermittently turn off.

Equipment: GE Healthcare - Panda iRES

RE: GE HealthCare - Panda iRES
Updated  Apr 7 2023
4 Replies

I have a Panda warmer that the bed will not go up or down.  I have tried the pedals on both sides, still no response.  However, I when I press the pedals, I do hear a relay clicking.  Please help. 

Equipment: GE Healthcare - Panda iRES

RE: GE HealthCare - Panda iRES
Updated  Jan 17 2023
6 Replies

When the Panda is powered on the display is dark and unreadable. It takes about 5 minutes for it to warm up and become useable. Can you replace just the backlight or does the entire display need to be replaced?

RE: GE HealthCare - Panda iRES
Updated  Jan 9 2023
3 Replies

I cannot get my IRES unit to pass the Supply Gas Check Valve test, it has been a while since I last tested this unit. I vaguely remember that there was a tube you have to clamp off inside the compartment to get this to pass, can anyone confirm this and tell me which hose should be clamped.

RE: GE HealthCare - Panda iRES
Updated  Jul 28 2022
1 Reply

What should be the first thing to be done in troubleshooting this? Compressed air and O2 supply connected.

Equipment: GE Healthcare - Panda iRES

RE: GE HealthCare - Panda iRES
Updated  Apr 9 2021
3 Replies

Do anyone know the part number for the Panda Warmer scale cable?

RE: GE HealthCare - Panda iRES
Updated  Nov 12 2020
1 Reply

When I turn on the machine in pre-warming everything is ok bur in like 10 min working when the system is already warm  the observations lights suddenly shut off and I can not turn in on again until the unit is cool again.