Hi. I know the X2 can be used with a CO2 module when connected to power.
Does anybody knows if the CO2 can be used if I connect an external battery as M4605A with a 865297 battery extender module? If it needs a special configuration or something special to achieve this?
I have been trying to do this but the X2 apparently doesn’t recognize the battery extender.
RE: Philips - IntelliVue X2Updated Dec 4 2023
Updated Nov 11 2019
1 ReplyHas anyone ever had any luck finding out what the codes are in the Status Log? If they're like the ones for the old Merlin CMS monitors, for instance, here's one I'd like to find the answer to:
H 16400 300 08 NOV 19 0200
Okay. As near as I can tell, H is for Hardware, 16400 is the generic code from Philips for the main hardware card, and.. 300 is an Operating System error, with the 300 in the Merlin being "EPROM checksum incorrect".
Any ideas? RE: Philips - IntelliVue X2
Updated Sep 1 2018
1 Replyhow can I get all 12 leads to show up when using the 5+5 leads ECG set on the 12 lead ecg profile?
I can get the chest lead of the Va & Vb selected in setup ECG displayed but not the other 4. Am I missing something with the setup?
Using Intellivue X2 attached to MX700 with M1663A trunk cable and M1968A limb leads, M1976A chest lead RE: Philips - IntelliVue X2
Updated Dec 14 2017
8 RepliesCan anyone recommend a company to repair NBP pumps for Philips MP/MX/X2 patient monitors; or FM 20/30/40/50 fetal monitors? RE: Philips - IntelliVue X2
Added Nov 1 2017
Updated Jan 3 2017
Updated Oct 11 2016
Updated Aug 15 2016
2 Replies
Staff keeps dropping these X2s and damaging the screens. Anyone have any thoughts on a screen protector maybe like something used on phones?
RE: Philips - IntelliVue X2