Hello please i need help when I press reports msg show option not available any idea?

RE: GE HealthCare - Logiq S8
Updated  Feb 19 2021
1 Reply

How do I setup volume measurement in order to do it both ways: start the measurement when pressing volume or take 3 calliper distances and press volume in the end? The last one is not working for me. THANKS.

RE: GE HealthCare - Logiq S8
Updated  Feb 15 2021
1 Reply

I have a logiq S8 that is having intermittent artifacts all over the scanning screen. Ran diagnostics for the noise flor and BF in low, med and high. Failed the noise test, channels 191 failed on low, med and high. I replaced the BF board(it wasn't cheap), a month later the issue comes back, it didn't fail the noise floor test but it did fail the same channels as before, plus channels 195 and 196 on the BF med test. Replaced the PID board and now we are seeing the same issue, has anyone seen this before?

RE: GE HealthCare - Logiq S8
Updated  Sep 6 2019
4 Replies

Hi, i am trying to connect a pencil cw-probe that i bought to my s8. The slot is blocked. I already did a system update with the cw key but i cannot physically connect. I suspect the board is not able to receive cw and i have to buy a new piece of hardware, something like a port. Do i have to buy a new main board or do you know of the existance of such a connector port for CW? Thank you in advance

RE: GE HealthCare - Logiq S8
2 Replies

Hello, does anyone have \WINDOWS\SYSTEM32\CONFIG\SYSTEM file or complete disk image? Thanks!

RE: GE HealthCare - Logiq S8
1 Reply

How do you turn off the harmonics on the machine

RE: GE HealthCare - Logiq S8
Updated  Oct 24 2017
2 Replies

Does anyone have a copy of the service manual for the Logiq S8? Thanks in advance.

RE: GE HealthCare - Logiq S8
Updated  Oct 24 2017
1 Reply

Does anyone have a copy of the service manual for the Logiq S8? Thanks in advance.

RE: GE HealthCare - Logiq S8
Updated  Oct 24 2017
1 Reply

Hello, I have a Logiq S8 that when we send images to our PACs the still images have the patient information displayed in banner at the top of the image. However the clips do not have the banner or any patient information displayed. Does anyone know where to configure to turn that on so we get the patient info on our clips. thanks

RE: GE HealthCare - Logiq S8
Updated  Oct 24 2017
3 Replies

Does anyone have a copy of the service manual for the Logiq S8? Thanks in advance.